OxiTop
®
Control OC 100
What to do if...
ba31114e06
06/2012
69
Memory lack!
No ready samples
autom. erasable!
The instrument works with the set-
ting, Erase memory
auto
(automati-
cally). The memory is full and only
unfinished
samples are stored in the sample
management.
The instrument can only erase fin-
ished samples
automatically.
Transfer the unfinished samples of
the sample management to
“finished” status.
To do this, the measurement data
of all measuring heads must
be called up and evaluation and
documentation of the measurement
data of the finished samples must
be performed.
If the measurement data of the
measuring heads are not yet fin-
ished, no instant remedy is possi-
ble!
The memory capacity limit of the
instrument has been reached.
Memory lack!
Erase finished
sample/s!
The instrument works with the set-
ting, Erase memory "manually". The
memory is full.
Transfer the unfinished samples of
the sample management to
“finished” status:
Call up the measurement data of all
measuring heads
Perform the evaluation and
documentation for the measure-
ment data of the finished
samples
Erase finished samples from the
sample management:
section 7.6.1 E
Note: Only the finished samples in the
sample management can be erased.
Display message
Cause
Remedy
Содержание OxiTop C
Страница 6: ...Contents OxiTop Control OC 100 6 ba31114e06 06 2012...
Страница 14: ...Commissioning OxiTop Control OC 100 14 ba31114e06 06 2012...
Страница 55: ...OxiTop Control OC 100 GLP TOOLS ba31114e06 06 2012 55...
Страница 60: ...Transmitting data OxiTop Control OC 100 60 ba31114e06 06 2012...
Страница 80: ...Index OxiTop Control OC 100 80 ba31114e06 06 2012...
Страница 81: ......