OxiTop
®
Control OC 100
GLP/TOOLS
ba31114e06
06/2012
43
7.4.5
GLP
- calibration interval (
- Calinterval
)
Set the time period (1 to 36 months) here. When it expires, the instrument regis-
ters that the next test resource monitoring of the measuring heads is due. After
the calibrating interval expires, measuring is blocked until the calibration is
performed or the
Off
setting is selected.
7.4.6 Data storage
Here, you can set whether the controller should automatically erase the oldest
finished samples if the memory is full to create space for new measurement
data (setting
auto
).
With the
manual
setting, the message "
Memory lack!
Erase finished
sample/s!
is displayed if the memory is full (see section 10 W
GLP
menu:
Preselected:
- GLP
Confirm with
<RUN/ENTER>
.
Use
<
><
>
to select either
GLP
On
or
Off
.
Confirm with
<RUN/ENTER>
.
If the end of the calibrating interval set up lies within the measuring
time of a measurement to be started, this measurement cannot be
started.
GLP
menu:
Select the
- Calinterval
menu item
with
<
><
>
Confirm with
<RUN/ENTER>
.
Use
<
><
>
to set the time
period for the calibrating interval (1
- 36 months).
Delivery setting:
12 months.
Confirm with
<RUN/ENTER>
.
Back up your data regularly!
RUN/ENTER
RUN/ENTER
Содержание OxiTop C
Страница 6: ...Contents OxiTop Control OC 100 6 ba31114e06 06 2012...
Страница 14: ...Commissioning OxiTop Control OC 100 14 ba31114e06 06 2012...
Страница 55: ...OxiTop Control OC 100 GLP TOOLS ba31114e06 06 2012 55...
Страница 60: ...Transmitting data OxiTop Control OC 100 60 ba31114e06 06 2012...
Страница 80: ...Index OxiTop Control OC 100 80 ba31114e06 06 2012...
Страница 81: ......