OxiTop
®
Control OC 100
Technical data
ba31114e06
06/2012
77
11 Technical data
11.1 Controller OxiTop
®
OC 100
Measuring ranges
0 ... 40 / 80 / 200 / 400 / 800 / 2000 / 4000 mg/l BOD
Run durations and data
records
Measuring time of the
measurement
Data records
Per measurement
0.5 hours
180
1.5 hours
270
2.5 hours
300
3.5 hours
315
4.5 hours
324
5.5 hours
330
All whole hours
(1, 2, 3 to 23 hours)
360
All whole days
(1, 2, 3 to 99 days)
360
Display
LCD graphic display 64 x 128 pixels
Memory capacity
Data of 120 measuring heads
Power supply
Supply batteries: 3 pieces, alkaline (alkali manganese), size:
Mignon, AA, AM3, LR6
Data backup battery: 1 piece lithium battery, CR2430, WTW
order no. 209 012
Battery runtime
Supply battery
Data backup battery
> 1000 start-ups in case of normal use
(the instrument has an automatic shut off)
typically 4 years
Protective class
3, instrument safety: IEC 1010, EN 61010 part 1
Type of protection
IP 41 EN 50529
EMC
-Interference emission
-Interference immunity
EN 50081-1, FCC Class A
EN 50082-1
Climatic class
2, VDI/VDE 3540
Ambient temperature
Storage: -25°C ... +65°C
Operation: +5°C ... +40°C
Dimensions
45 x 100 x 200 mm (H x W x D)
Weight
Approx. 390 g
Test certificates
CE, UL, CUL(UL/CSA)
Содержание OxiTop C
Страница 6: ...Contents OxiTop Control OC 100 6 ba31114e06 06 2012...
Страница 14: ...Commissioning OxiTop Control OC 100 14 ba31114e06 06 2012...
Страница 55: ...OxiTop Control OC 100 GLP TOOLS ba31114e06 06 2012 55...
Страница 60: ...Transmitting data OxiTop Control OC 100 60 ba31114e06 06 2012...
Страница 80: ...Index OxiTop Control OC 100 80 ba31114e06 06 2012...
Страница 81: ......