7.5.3 Controller information
7.5.4
cal-test
The
cal-test
function is used to test the sealing of the system measuring head
- rubber sleeve - sample bottle and the operability of the system OxiTop
®
Control.
In the
BOD Standard
mode with the
GLP
On
setting, the controller indicates
when the next cal test is due, depending on the set calibration interval (see
section 7.4.4 GLP (BOD S
To perform the test, you need the WTW test resource, OxiTop
®
PM, order
number 209 333.
Sample preparation: see operating manual of the OxiTop
®
PM test resource.
CHECK
menu.
Select the
- Controller Info
menu
item with
<
><
>
.
The display shows the following
information:
Number of measuring heads
for
which there is still memory
capacity
Status of the supply batteries
Status of the data backup
batteries (
OK
/
LOBAT
).
Use
<
><
>
to scroll to the
specification of the
Software version
Series number
(can differ from the example
shown here)
Return to the
CHECK
menu:
With
<RUN/ENTER>
.
CHECK
menu.
Select the
- Cal Test
menu item
with
<
><
>
.
RUN/ENTER