Operation
System 182-XT-4
4 - 16
ba75753e06
12/2014
Putting the sensor into
operation after
calibrating
5
Confirm the calibration data with
g
. Calibration is completed
with this. The following display message describes the further
steps to put the sensor into operation again.
6
Confirm with
g
. The display returns to the measured value
display (the measured value flashes as the sensor is still in the
maintenance condition).
7
If calibration was successful, immerse the sensor in the test
sample.
8
Wait for a stable measured value.
9
Switch off the maintenance condition (highlight sensor and
press
g
three times). The measured value has stopped
flashing.
If calibration was not successful, this is indicated by "----"
on the display and a corresponding message with remedial
actions appears in the log book. Follow the instructions and
repeat calibration.