Technical data
System 182-XT-4
9 - 4
ba75753e06
12/2014
Relay
(6 x)
Current outputs
(5 x)
Output
Galvanically separated
Max. switching voltage
Front terminal strip:
240 VAC or 24 VDC
Rear terminal strip:
250 VAC or 24 VDC
Max. switching current
Front terminal strip:
2 A (AC and DC)
Rear terminal strip:
5 A (AC and DC)
Fuse rating on the
operator side
Front terminal strip:
max. 2 A
Rear terminal strip:
Maximum 5 A
Relay functions
Programmable as:
– Opener or closer
– Limit monitor
– Monitoring of the warning and error
signals of the DIQ/S 182-XT-4
– Proportional frequency output
– Proportional pulse width output
Output
Galvanically separated from the sensors
Output current
Can be switched between 0 - 20 mA and
4 - 20 mA
In the case of errors, can be set to: 0 ... 21
mA
Max. initial output
voltage
24 V, in the case of missing or incorrect
burden
Accuracy
0.3 % of current value ± 50 µA,
load max. 500
Ω
Functions
Programmable:
– Recorder attenuation, adjustable
0 - 40 mA/s
– Error behavior can be set as required
0 ... 21 mA
– Error behavior according to Namur NE43
can be selected
– Positive and negative characteristic
curve
– PID controller
Содержание DIQ/S 182-XT-4
Страница 6: ...Contents System 182 XT 4 0 4 ba75753e06 12 2014 ...
Страница 14: ...Safety System 182 XT 4 2 4 ba75753e06 12 2014 ...
Страница 68: ...Installation System 182 XT 4 3 54 ba75753e06 12 2014 ...
Страница 86: ...Operation System 182 XT 4 4 18 ba75753e06 12 2014 ...
Страница 92: ...The SETTINGS menu System 182 XT 4 5 6 ba75753e06 12 2014 ...
Страница 128: ...Maintenance and cleaning System 182 XT 4 7 2 ba75753e06 12 2014 ...
Страница 144: ...Technical data System 182 XT 4 9 12 ba75753e06 12 2014 ...
Страница 146: ...Accessories and options System 182 XT 4 10 2 ba75753e06 12 2014 ...
Страница 150: ...Appendix store separately if required System 182 XT 4 12 2 ba75753e06 12 2014 ...
Страница 151: ......