Fig. 2 — Initial syringe clamp adjustment —
Secure syringe by sequentially
adjusting set screws with supplied Allen key:
1
for horizontal alignment,
2
for
vertical alignment, and
3
to eliminate play.
Axial Needle Alignment
in order to maintain a good syringe needle alignment particularly along the same
axis of the supporting bar, rotate the syringe body while placing in the two clamps
to allow the syringe to seat properly and align along the body of the pump for
minimal slant offset. if the clamp which holds the syringe collar (see “collar stop”
in the drawing below) is too tight or too loose, syringe needle alignment may be off
because of stress.
if the collar stop is too tight to allow the syringe collar to insert easily, then follow
these instructions:
immediately below and behind the collar stop, in the groove where the long drive
screw is, there is a small Phillips head screw, attached to the collar stop. loosen
this screw to slightly (about 0.5 to 1 mm) to allow for a thicker collar (see Fig 2).
You may have to grasp the collar stop and wiggle it backwards to move it. once
the stop is backed out; check and adjust for a tight fit so the syringe body doesn’t
move when placed into the holder. Gently re-tighten the screw in the new position.
Note:
you may have to retract the plunger button holder to get access to the set-
screw.
Mount the pump
the ultramicroPump can
be mounted directly onto
a stereotaxic frame or a
micromanipulator, using
the mounting bar (see
Fig. 2). the mounting bar
may be unscrewed and
modified by the user, if
necessary.
MOUNTING BAR
COLLAR
STOP
SYRINGE
COLLAR
SET SCREW
(not visible in this view)
PLUNGER
RETAINING
SCREW
PLUNGER BUTTON
PLUNGER
BUTTON
HOLDER
CABLE CONNECTS
TO REAR PANEL OF
MICRO4 CONTROLLER
SYRINGE
CLAMPS
CLAMP
RELEASE
SET
SCREWS
SYRINGE
1
2
3