UltraMicroPump II
World Precision instruments
2
operating parameters for the ultramicroPump are set with the micro4. up to four
pumps may be independently controlled. operating parameters set by the user are
stored in “non-volatile” memory for instant recall when the unit is powered on.
An optional footswitch can be plugged into an rs232 port on the rear of the
controller for “hands free” start / stop operation. the same port may also be used to
connect the controller to a computer or to some other device for ttl triggering.
Parts List
After unpacking, make certain that there is no visible damage to the instrument.
check to see that all items are included:
UMP2
ultramicroPump ii
UMC4
micro4
™
controller, 4-channel
40300
12-volt Power supply
Power cable
Allen key
instruction manual
For a list of microsyringes available from WPi, see page 23.
Set-up and Operation
these instructions will help you put the ultramicroPump to immediate use. We
recommend that you read the entire manual and familiarize yourself with the
various operating procedures of the ultramicroPump and micro4.
Mounting the syringe
syringes may be filled manually before mounting in the ultramicroPump or filled by
using the fast reverse function (see page 13). Place the plunger button of the filled
syringe into the plunger button holder (leaving the plunger retaining screw loose)
while also placing the syringe collar into the collar stop (see Fig. 2). take care not
to damage the syringe collar during this installation. Gently tighten the plunger
retaining screw so that the plunger button will not move independently when the
pump is activated (this allows for zero volume error during pump operation).