UltraMicroPump II
World Precision instruments
22
General Maintenance
ultramicroPump requires minimal maintenance; regular laboratory cleaning will
keep this instrument in optimum operating condition.
do not apply solvents or oils to any part of the ultramicroPump.
this instrument is
not autoclavable
.
do not disassemble—there are no serviceable parts inside either the umPii or the
micro 4 controller.
Always hold ultramicroPump by the main body or mounting bar. do not swing or
carry the ultramicroPump by its cable.
Storage:
store the umPii in a sealed plastic zip bag to prevent dust from accumulating on
the drive screw. excessive dust can cause jams and inadvertent stops.
use of gas-tight syringes on the umPii is not recommended for syringes above
250 µl as this can damage the motor. Please use liquid-tight syringes for
applications that require volumes greater than 250 µl.
Troubleshooting
Problem:
the instrument motor makes noise but the plunger button does not travel.
Check:
look for a loose connector at the rear of the micro4, make sure the umPii
plug is firmly seated. the gray plastic plug should be a flush fit with the connector
on the controller. Verify that the pins in the connector are not damaged.
Check:
test the pump in another channel, with the same program parameters.
Instrument Overtravel
if the plunger button has traveled to the extreme edge of the pump and has
jammed, try the following procedure to free the plunger button:
Place the pump so that the syringe would be pointing to the right (no syringe
should be in the pump). Program the micro4 to syringe style F (or larger to H),
program in about 2000 to 5000 nl in the volume and program in a rate of 2000 or
greater (there are internal limits). Press and hold the right or left arrow key for the