5.
If the electrode is not clean, repeat steps 3 and 4. If necessary the ISO-NOP
Rejuvenator (WPI #
JUV
) can be used restore sensitivity of an old electrode
(contact WPI for assistance).
6.
Remove the locking cap from the old used sleeve, and gently slide it onto the
new replacement sleeve.
7.
Dip the internal electrode 1-2 cm into the electrolyte provided in the ISO-NO
start-up kit, the current should go offscale during this. Using the MicroFil™
nonmetallic syringe needle and 1 mL plastic syringe, supplied, inject
approximately 100 microliters of electrolyte directly into the new sleeve. The
MicroFil supplied should be less than the length of the sleeve, so that it will not
puncture the delicate membrane at the tip of the sleeve during injection. If the
MicroFil is longer than the sleeve it can be cut to the correct length.
8.
Slowly and smoothly insert the electrode into the sleeve, and screw the locking
cap into the handle. The electrode should be observed to press gently against
the membrane.
9.
The current displayed on the meter at this time will be high or offscale.
10.
Suspend the tip of the new assembled probe in distilled water.
11.
After 10-15 minutes the current should no longer be offscale and will gradually
decrease with time. It may take several hours for the sensor current to reach a
low stable value, at which time it will be ready for use.
12.
The integrity of the new membrane can be determined by immersing the probe
tip into a strong saline solution (1 M). If the current observed, after a few
minutes in the saline solution, increases dramatically or is offscale then the
membrane integrity is not good and a new membrane will have to be fitted.
13.
When the ISO-NOP is not being used it should be stored with the tip
suspended in distilled water.
Additional membrane kits (WPI #
5436
) may be purchased separately.
Maintenance of Nitric Oxide Microsensors
WPI’s Nitric Oxide microsensors are maintenance-free consumable sensors that
are warranted against defect for 30 days from the date of purchase. The following
information should increase the lifetime of the sensor:
Содержание APOLLO 4000
Страница 2: ......
Страница 4: ......
Страница 10: ...APOLLO 4000 INTRODUCTION WORLD PRECISION INSTRUMENTS A 6...
Страница 48: ...APOLLO 4000 NITRIC OXIDE DETECTION WORLD PRECISION INSTRUMENTS C 24...
Страница 70: ...APOLLO 4000 TROUBLESHOOTING WORLD PRECISION INSTRUMENTS F 2...
Страница 72: ......
Страница 74: ......