APOLLO 4000
OPERATING INSTRUCTIONS
WORLD PRECISION INSTRUMENTS
B-3
NOTE:
The Temp and Sensor inputs are not interchangeable but no damage
to the APOLLO 4000 will occur if a sensor/electrode is accidentally inserted
into the wrong input receptacle.
To connect a sensor, simply line up the red dot on the metal connector attached to
the sensor cable with the red dot on the sensor input receptacle and insert the
cable connector (
Fig. B1
). An LED (located above each sensor input) will light up
immediately indicating the instrument and sensor are connected and working
correctly. The Temp LED is red. The Sensor LED is green. The following LED
indications will inform the user of the status of the system:
LED CONDITION
INDICATION
Temp LED (red)
Steady RED light
Electrode is performing normally
No RED light
Electrode is not connected or is damaged
Sensor LED (green)
Steady GREEN light
Sensor is performing normally within the
user-selected current range
Intermittent blinking GREEN light
Sensor current is outside the user-
selected linear range.
No GREEN light
ERROR sensor current range too low.
Sensor error or sensor not connected.
NOTE:
WPI strongly recommends that APOLLO 4000 be powered
through a Back-UPS unit to avoid system failure during power loss.
It is the responsibility of the user as well to install appropriate anti-
virus protection software.
WPI will not be liable for any loss of data as a result of power loss or virus-attack to the APOLLO 4000 system.
Содержание APOLLO 4000
Страница 2: ......
Страница 4: ......
Страница 10: ...APOLLO 4000 INTRODUCTION WORLD PRECISION INSTRUMENTS A 6...
Страница 48: ...APOLLO 4000 NITRIC OXIDE DETECTION WORLD PRECISION INSTRUMENTS C 24...
Страница 70: ...APOLLO 4000 TROUBLESHOOTING WORLD PRECISION INSTRUMENTS F 2...
Страница 72: ......
Страница 74: ......