www.wobit.com.pl 

User manual  -ADT42 – 23.06.2015 – v.1.3 

   page 6 

 

 

2.4

 

Power supply 

ADT42 module can be powered by voltage in range 12 - 24 VDC (max. 36 VDC). Current consumption 
is max. 150mA (for connected 4 force sensors). While using transistor outputs you should add current 
consumed on this outputs (max. 0,5A / output). 
 

2.5

 

Force sensors connection

 

ADT42  module  can  cooperate  with  any  force  sensor  which  has  bridge  with  resistance  bigger  than 
150Ω. Sensor(s) should connect according to scheme below. 

 

 

Picture. 2 Connection of force sensors. 

 

 

CAUTION!

 

Sensor wires usually have additional shielded braid, which is connected with metal housing of the 
sensor. This signal can be connected to supply ground of ADT42, but we don’t recommend to do 
that, while sensor is mounted to metal construction, which can be in different potential. 

 

 

 

CAUTION!  

Do not short-circuit derivation cables of sensor power supply (E+,E-), and also do not short-circuit 
E+ output to other  derivation, it can cause damage of the device. 

 

Sensor model 

 

Signals 

K200, K300, K500, 
KB52, KB82, KMB19, 
KMB25, KMB31 

K701, K801, 
K1101, K1506 

K1401, K1501, 
K1600 

KMM20, KMM30, KMM40, 
KMM50, KMM60, K1505, 
EMS70, EMS150, EMS200 

Power  supply  + 
(V+) 

Red 

Brown 

Red 

White 

Power  supply  - 
(V-) 

Black

 

Yellow 

Black

 

Brown 

 (S+) 

Green 

Green 

Brown 

Yellow 

Signal - (S-) 

White 

White 

Yellow 

Green 

Содержание ADT42

Страница 1: ...s c 62 045 Pniewy D borzyce 16 tel 061 22 27 422 fax 061 22 27 439 e mail wobit wobit com pl www wobit com pl ADT42 User Manual 4 channel measuring module for force sensors with 0 10V 4 20mA output U...

Страница 2: ...MODBUS Communication 12 5 Complementary information 14 5 1 Force sensor types 14 5 2 Sensor measuring range 14 6 Technical parameters 15 Thank you for selecting our product This instruction will help...

Страница 3: ...the device or to fire Dismantling of the indicator housing during guarantee agreement period results in its invalidation 1 2 Assembly recommendation In the environments of unknown levels of interrupti...

Страница 4: ...of measurements from selected channels the module can be used in multi sensor weight systems ADT42 has 2 transistor outputs and 2 insulated inputs one tare input Built in operating modes enable the de...

Страница 5: ...surement data acquisition 2 3 Description of connectors and diodes Picture 1 Connector and panel description Nr Description Analog output 0 10V 4 20mA 1 GND Ground of analog output 2 AOUT 0 10V 4 20mA...

Страница 6: ...res usually have additional shielded braid which is connected with metal housing of the sensor This signal can be connected to supply ground of ADT42 but we don t recommend to do that while sensor is...

Страница 7: ...N2 0 output signal aggregate signal after tare IN2 1 output signal aggregate signal before tare Tab 2 Functions of opt insulated inputs Picture 3 Opt insulated input 2 7 Transistor outputs ADT42 has t...

Страница 8: ...s the module Good quality USB cable with length 1 5 m equipped with ferrite beads Opt insulated USB HUBs on PC side In some conditions can occur that communication won t be possible or measurement dat...

Страница 9: ...or using option Reverse the sign and zeroing channel by pressing Zero Filtration of measurement ADT42 has constant averaging filter set in range 0 99 samples and measurement mode Fast and Slow 2 Bigge...

Страница 10: ...g output A selection of signal source sent on 0 10V 4 20mA output B calibration of analog output 2 Setting of transistor outputs OUT 1 and OUT2 3 Setting of communication parameters RS485 MODBUS 4 Rea...

Страница 11: ...3 Set Output 0V for value 500 4 Set Output 10V for value 0 5 Save settings by Save button 3 3 2 Configuration of transistor outputs By using this option it is possible to configure conditions of acti...

Страница 12: ...5 CANAL1_INT INT R 0x03 Value of 2 channel 5 6 CANAL2_INT INT R 0x03 Value of 3 channel 6 7 CANAL3_INT INT R 0x03 Value of 4 channel 7 8 ANALOG_INT INT R 0x03 Value of analog output x1000 Values store...

Страница 13: ...tion table Reading of measurements from SUM_IREAL register Function 03 Register address 0 Request MODBUS MASTER ADT42 Response ADT42 MODBUS MASTER Device address 0x01 Device address 0x01 Function 0x03...

Страница 14: ...measured force 0 200 kN Example of models K1101 5 2 Sensor measuring range Rated load for sensor it is a size of force which specifies top value of measuring range given value for sensor as bridge ran...

Страница 15: ...e error 0 0025 C Measurement frequency filtration mode slow fast 1 active channel 10 Hz 80 Hz 2 active channel 1 2 Hz 10 Hz 3 active channel 0 8 Hz 6 5 Hz 4 active channel 0 6 Hz 5 Hz Time of initial...

Отзывы: