OPERATION
POWER LEVEL ACCURACY
VERIFICATION
AND FLATNESS TESTS
Model 68185B
Serial No. ____________________
Date ___________
Model 68185B
(with Option 2D Step Attenuator)
Power Level Accuracy *
(CW Frequency = 5.0 GHz)
Power Level Accuracy *
(CW Frequency = 25.0 GHz)
Power Level Accuracy *
(CW Frequency = 50.0 GHz)
Set Power
Measured Power
Set Power
Measured Power
Set Power
Measured Power
+ 8.5 dBm
+ 7.5 dBm
+ 6.5 dBm
+ 5.5 dBm
+ 4.5 dBm
+ 3.5 dBm
+ 2.5 dBm
+ 1.5 dBm
+ 0.5 dBm
– 0.5 dBm
– 1.5 dBm
– 2.5 dBm
– 3.5 dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
+ 0 dBm
– 1 dBm
– 2 dBm
– 3 dBm
– 4 dBm
– 5 dBm
– 6 dBm
– 7 dBm
– 8 dBm
– 9 dBm
–10 dBm
–11 dBm
–12 dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
– 1.5 dBm
– 2.5 dBm
– 3.5 dBm
– 4.5 dBm
– 5.5 dBm
– 6.5 dBm
– 7.5 dBm
– 8.5 dBm
– 9.5 dBm
–10.5 dBm
–11.5 dBm
–12.5 dBm
–13.5 dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
* Specification is
±
1.0 dB.
* Specification is
±
1.0 dB.
* Specification is
±
1.5 dB.
Power Level Flatness (Step Sweep)
Set Power
Max Power
Min Power
Variation **
– 2 dBm
_________dBm
__________dBm
_________dB
** Maximum variation is 1.6 dB (0.5 to 40 GHz); 2.2 dB (40 to 60 GHz).
Power Level Flatness (Analog Sweep)
Set Power
Max Power
Min Power
Variation ***
– 2 dBm
_________dBm
__________dBm
_________dB
*** Maximum variation is 6.0 dB (0.5 to 20 GHz); 8.2 dB (20 to 40 GHz); 10.2 dB (40 to 60 GHz)(typical, not a
specification).
Table 5-3.
Power Level Accuracy and Flatness Test Record (42 of 46)
5-60
681XXB OM