At the Configure GPIB menu, press
More
to go to
an additional Configure GPIB menu (below).
Press
Pwr Mtr Address
to change the address of
the power meter on the GPIB (the power meter’s de-
fault address is 13). Enter the new address, be-
tween 1 and 30, using the cursor control key or the
data entry key pad and the terminator key
Hz
ns
ADRS
The new GPIB address will appear on the display.
Press
Pwr Mtr Select
to select the power meter
model being used. (Supported power meters are the
Anritsu ML4803A and Hewlett-Packard 437B,
438A, and 70100A.)
Press
Previous Menu
to return to the main Config-
ure GPIB menu display.
At the Configure GPIB menu, press
Previous Menu
to return to the System Configuration menu display.
Creating a Power-Offset Table
The 681XXB must be in CW frequency mode and
fixed (non-swept) power level mode in order to cre-
ate a power-offset table for user level flatness correc-
tion.
Place the sweep generator in CW frequency mode
by pressing the main menu key
CW/SWEEP
SELECT
At the resulting menu display, press
CW
. The
681XXB is now in CW frequency mode.
LOCAL (FRONT
LEVELING
PANEL) OPERATION
OPERATIONS
681XXB OM
3-57