OPERATION
POWER LEVEL ACCURACY
VERIFICATION
AND FLATNESS TESTS
Model 68147B w/Option 15A
Serial No. ____________________
Date ___________
Model 68147B with Option 15A High Power
(with Option 2A Step Attenuator)
Power Level Accuracy *
(CW Frequency = 1.0 GHz)
Power Level Accuracy *
(CW Frequency = 5.0 GHz)
Set Power
Measured Power
Set Power
Measured Power
+11 dBm
+10 dBm
+ 9 dBm
+ 8 dBm
+ 7 dBm
+ 6 dBm
+ 5 dBm
+ 4 dBm
+ 3 dBm
+ 2 dBm
+ 1 dBm
+ 0 dBm
– 1 dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
+15 dBm
+14 dBm
+13 dBm
+12 dBm
+11 dBm
+10 dBm
+ 9 dBm
+ 8 dBm
+ 7 dBm
+ 6 dBm
+ 5 dBm
+ 4 dBm
+ 3 dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
* Specification is
±
1.0 dB.
* Specification is
±
1.0 dB.
Power Level Flatness (Step Sweep)
Set Power
Max Power
Min Power
Variation **
+11 dBm
_________dBm
__________dBm
_________dB
** Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 1.6 dB (0.05 to 20.0 GHz).
Power Level Flatness (Analog Sweep)
Set Power
Max Power
Min Power
Variation ***
+11 dBm
_________dBm
__________dBm
_________dB
*** Maximum variation is 7.0 dB (0.0-1 to 0.05 GHz); 6.0 dB (0.05 to 20.0 GHz)(typical, not a specification).
Table 5-3.
Power Level Accuracy and Flatness Test Record (12 of 46)
5-30
681XXB OM