10. Specifications
Indicator model CTH6200
Alarm
Min./Max. alarm (audible/visual); can be set via menu
Min./Max. memory
Minimum or maximum measured value; can be accessed via
function button
Sensor characteristic
curve
Standardised: E.751
Characteristic curve per EN 60751
Application-specific
sensor characteristic
curve
■
Up to 50 programmable points
■
Adjustable via separate software
Data logger
Individual value logger
Up to 99 recordings incl. time can be
accessed via function button
Cyclic logger
■
Automatic recording of up to 16,384
values incl. time
■
Cycle time freely adjustable in the
range from 1 ... 3,600 seconds
Power-off function
Automatic switch-off; can be set via menu
Activated
1 ... 120 minutes
Deactivated
No automatic switch-off of the
instrument
Voltage supply
Supply voltage
■
9 V battery
■
9 V rechargeable battery
■
Mains power supply
Battery life
Approx. 20 hours of operation
Current consumption
Approx. 1 mA, approx. 300 h
Permissible ambient conditions
Operating temperature
-25 ... +50 °C [-13 ... +122 °F]
Storage temperature
-25 ... +70 °C [-13 ... +158 °F]
Relative humidity
0 ... 95 % r. h. (non-condensing)
Output signals/interfaces
Connection
Stereo jack connector, 3.5 mm
Optional serial interface or analogue output
Serial interface
■
RS-232
■
USB
Instrument-specific interface cable required
Analogue output
■
DC 0 ... 1 V; configurable
■
Selectable via menu alternative to serial interface