26
WIKA operating instructions, model CTH6200
EN
2146872.02 07/2021 EN/DE
6. Commissioning, operation
Loads up to approx. 10 kΩ are harmless.
⇒
If the display exceeds the value set with
dAC
.
1
, then 1 V will be output.
⇒
If the display drops below the value set with
dAC
.
0
, then 0 V will be output.
⇒
In the event of an error (
Err
.
1
,
Err
.
2
, ----, etc.) an analogue signal slightly over 1 V will
be output.
Jack connector wiring
WARNING!
Damage to property through incorrect jack connector
Using an incorrect jack connector or incorrect wiring can cause damage to
the hand-held.
▶
The 3rd connection must not be used.
▶
Only stereo jack connectors are permitted.
▶
Only use the original connection cable from WIKA.
6.3.10 Alarm (
AL
.
)
Three settings are possible:
“
AL
.
oFF
”
= off
“
AL
.
on
”
= on
“
AL
.
no
.
So
”
= on without sound
Under the following conditions, an alarm is given when the alarm function
“
AL
.
on
”
or
“
AL
.
no
.
So
”
is active:
■
Value is below lower alarm limit
“
AL
.
Lo
”
or above upper alarm limit
“
AL
.
Hi
”
.
■
Sensor error (
Sens-Erro
)
■
Low battery,
'
bAt
'
■
Err
.
7
: System error (always indicated with a sound)
In the event of an alarm, the display flashes and the instrument beeps. If the serial
interface is used, the 'PRIO' flag is used.
6.3.11 Real-time clock (
CLOC
)
The real-time clock is needed for the time allocation of the logger data. If required, check
the settings.
Checking the real-time clock
1. Press the
[SET/MENU]
button for 2 seconds.
⇒
The main menu
“
SEt
”
is accessed.