21
WIKA operating instructions, model CTH6200
EN
2146872.02 07/2021 EN/DE
6. Commissioning, operation
Menu Para-
meter
Values
Meaning
[MENU]
▶
▲
or
▼
read
Lo66
Read Logg: Read manual individual value logger
(Only if data are available, see chapter 6.4 “Operation of the logger function”)
SEt
ConF
Set Configuration: General settings
Unit
°C
All temperatures in degree Celsius
1)
°F
All temperatures in degree Fahrenheit
1)
rES
0.1
Resolution 0.1 °C
1)
0.01
Resolution 0.01 °C
1)
Auto
Resolution is selected automatically
1)
Lin
E.751
Characteristic curve per EN 60751
1)
USEr
User characteristic curve (default: EN 60751
1)
Changeable via GSoft evaluation software)
OFFS
-2.50 ... 2.50 °C
or
-4.50 ... 4.50 °F
The zero point of the measurement will be displaced by this
value to compensate for deviations of the probe and the
measuring instrument.
1)
oFF
Zero point correction is deactivated (= 0.0 °C)
1)
SCAL
-2.000 ... 2.000
The measuring slope will be changed by this factor [in
%] to compensate deviations of probe and measuring
instrument.
1)
oFF
Slope correction factor deactivated (= 0.000)
1)
t
.
au6
1 ... 30
Mean value filter, duration in seconds
1)
oFF
Mean value filter is deactivated
1)
P
.
oFF
1 ... 120
Auto power-off delay in minutes.
If no button is pressed and there is no data transfer via the
interface, the instrument will switch itself off automatically
after this interval.
oFF
Auto power-off function deactivated (continuous operation)
Ovt
oFF
No output function, lowest current consumption
SEr
Instrument output is serial interface
dAC
Instrument output is analogue output
Adr
.
01, 11 ... 91
Base address of the instrument for interface communication
(only if serial interface is selected)
dAC
.
0
-200 ... +850 °C /
-328 ... +1,562 °F
Zero point setting for
Out
=
dAC
: Temperature at which the
analogue output should output 0 V (only for
Out
=
dAC
)
dAC
.
1
-200 ... +850 °C /
-328 ... +1,562 °F
Slope setting for
Out
=
dAC
: Temperature at which the
analogue output should output 1 V (only for
Out
=
dAC
)
SEt
AL
.
Set Alarm: Alarm function configuration
2)
AL
.
on
Alarm sensor is on; is indicated acoustically
no
.
So
Alarm sensor is on; is not indicated acoustically
oFF
No alarm function
AL
.
Lo
-200 °C ...
AL
.
Hi
Min. alarm limit (not if
AL
.
oFF
)
AL
.
Hi
AL
.
Lo
... +850 °C
Max. alarm limit (not if
AL
.
oFF
)