32
10. Technical specifications
Type
AGC MicroVision Electroforming Unit
Operating voltage
230V~ / 50-60 Hz (Ref. No.75-21-0001)
(115V~ / 50-60 Hz)
(100V~ / 50-60 Hz)
Maximum power consumption:
approx. 240 VA
Maximum current input:
230 V / ~0.96 A
115 V / ~2.10 A
100 V / ~2.36 A
Fuses
230 V 2 x T 3.15 A; 115, 110 V 2 x T 5 A
Bath operating temperature:
65 °C accuracy ± 2 °C
Temperature sensors
PT1000 sensor
1 x external on heating ring
1 x on electroforming head
Safety temperature cut-out
Emergency shutdown if the maximum
heating temperature is exceeded
Bath volume
55 ml to 502 ml
200 ml glass beaker
55 ml to 150 ml
600 ml glass beaker
150 ml to 502 ml
Number of dies for electroforming
1-9 objects per electroforming process
Layering thicknesses
0.2 and 0.3 mm
(can be custom selected for each part)
Power levels
are determined with the aid of a comparison chart
Display type
5.7" LCD display with touch screen
Operation
Touch screen
Speed
350 - 550 RPM - ± 5 % (fixed)
Data input/output
for changing parameters and recording via a
USB interface (optional)
Dimensions:
210
mm wide,
300
mm deep,
450
mm high
Weight with glass beaker:
approx.
12 kg
incl. beaker
Layer thickness
Level
Bath volume
Weight
Layer thickness
Level
Bath volume
Weight
0.2 mm
1
2
3
4
5
6
7
9 ml
24 ml
41 ml
112 ml
159 ml
171 ml
188 ml
0.15 g
0.4 g
0.7 g
1.9 g
2.7 g
2.9 g
3.2 g
0.3 mm
1
2
3
4
5
6
7
12 ml
32 ml
56 ml
153 ml
218 ml
234 ml
258 ml
0.21 g
0.55 g
0.96 g
2.60 g
3.70 g
3.97 g
4.38 g
bar table