37
Pro Series
Furnace Operations Manual
0 SPEC – BURN IN
ENTRY TEMP 500C
Figure 56
0 SPEC – BURN IN
HEAT RATE 40C/M
Figure 57
0 SPEC – BURN IN
FINAL TEMP
Figure 58
0 SPEC – BURN IN
HOLD TIME 15:00
Figure 59
0 SPEC – BURN IN
LOWER TABLE 1000C
Figure 60
0 SPEC – BURN IN
COOL TIME 0:00
Figure 61
0 SPEC – BURN IN
USE vACUUM NO
Figure 62
SELECT PROGRAM?
0 SPEC – BURN IN
Figure 63
2. Muffle Burn In
DO NOT FORGET TO PLACE THE FIRING TRAY
ON THE LIFT TABLE!
A. Press the
Change Programs
button.
B. Turn the
Select Knob
right to select ADD A
PROGRAM. Press the
Next
button.
C. Select program 0. Press the
Next
button.
D. The furnace will ask for a porcelain type. Turn
the
Select Knob
to select SPEC. Press the
Next
button.
E. The furnace will tell you to enter a program name.
Turn the
Select Knob
and press the
Next
button after every letter. For blank spaces merely
press the enter button without turning the
Select
Knob
. Enter the name BURN_IN.
F. Enter an entry time of 1 minute by turning
the
Select Knob
. Press the
Next
button.
G. Turn the
Select Knob
to select an entry
temperature of 500C. Press the
Next
button.
H. The furnace will ask for a soak temperature,
press the
Next
button to skip this item.
I. The furnace will ask for a soak time,
press the
Next
button to skip this item.
J. Turn the
Select Knob
to select a heat rate of
40C/Minute. Press the
Next
button.
K. Turn the
Select Knob
to enter a final temperature
of 1000C. Press the
Next
button.
L. Turn the
Select Knob
to select a hold time of
15 minutes. Press the
Next
button.
M. The furnace will prompt you to enter an SP CL
TIME, press the
Next
button to skip.
N. Turn the
Select Knob
to select a temperature of
1000C to lower the table. Press the
Next
button.
O. The furnace will ask you to enter a cool time, press
the
Next
button to skip. This will automatically
enter a cool time of 0:00.
P. Turn the
Select Knob
to select NO vacuum.
Press the
Next
button.
Q. Press the
Run Program
button.
R. Turn the
Select Knob
to select Program
0 SPEC – BURN IN.
S. Press the
Special Functions
button.
T.
Turn the
Select Knob
to CALIBRATE OVEN.
Press the
Next
button.
U. Turn the
Select Knob
to the password number
260. Press the
Next
button.
The program will begin to run immediately and will cycle
until the
Stop Program
button is pressed. Leave the
furnace to cycle this program for eight hours. When
you return press the
Stop Program
button.
Fire a chip of porcelain to test the firing temperature of the
new muffle. Adjust the calibration as described in
Chapter
Four — Running a Program.
Adjust furnace calibration
to match your porcelain. Run Vacuum Test, see page 38.
15946 Pro Series English.indd 37
5/9/08 2:51:14 PM