13
WHILL Model F User Manual
1
1. Safety precautions
Table 3 - Proximity fields from RF wireless communications equipment
Test frequency
(MHz)
Band
(MHz)
Immunity test levels
Home healthcare environment
385
380-390
Pulse modulation 18 Hz, 27 V/m
450
430-470
FM, ±5 kHz deviation, 1 kHz sine, 28 V/m
710
704-787
Pulse modulation 217 Hz, 9 V/m
745
780
810
800-960
Pulse modulation 18 Hz, 28 V/m
870
930
1720
1700-1990
Pulse modulation 217 Hz, 28 V/m
1845
1970
2450
2400-2570
Pulse modulation 217 Hz, 28 V/m
5240
5100-5800
Pulse modulation 217 Hz, 9 V/m
5500
5785
Table 4 - Input a.c. power Port
Phenomenon
Basic EMC standard
Immunity test levels
Home healthcare environment
Electrical fast transients/
burst
IEC 61000-4-4
±1 kV
±2 kV
100 kHz repetition frequency
Surges
Line-to-line
IEC 61000-4-5
±0.5 kV, ±1 kV
Conducted disturbances
induced by RF fields
IEC 61000-4-6
3 V, 0.15 MHz-80 MHz
6 V in ISM bands and amateur radio bands
between 0.15 MHz and 80 MHz
80% AM at 1 kHz
Voltage dips
IEC 61000-4-11
0% U
T
; 0.5 cycle
At 0º, 45º, 90º, 135º, 180º, 225º, 270º and 315º
0% U
T
; 1 cycle
and
70% U
T
; 25/30 cycles
Single phase: at 0º
Voltage interruptions
IEC 61000-4-11
0% U
T
; 250/300 cycles
Содержание F
Страница 1: ...User Manual WHILL Model F User Manual en WHILL Model F ...
Страница 18: ...17 WHILL Model F User Manual 1 1 Safety precautions ...
Страница 56: ...55 WHILL Model F User Manual 3 3 Assembly and adjustment ...
Страница 92: ...91 WHILL Model F User Manual 6 6 Folding Procedure ...
Страница 100: ...99 WHILL Model F User Manual 7 7 Storage and transport methods ...
Страница 102: ...101 WHILL Model F User Manual 8 8 About the smartphone app ...
Страница 118: ......
Страница 119: ......
Страница 120: ...32 14016 1 01 2021 12 ...