EMI Compliance Table (Table 1)
Table 1 - Emission
Phenomenon
Compliance
Electromagnetic environment
RF emissions
CISPR 11
Group 1, Class B
Home healthcare environment
Harmonic distortion
IEC 61000-3-2
Class A
Home healthcare environment
Voltage fluctuations and
flicker
IEC 61000-3-3
Compliance
Home healthcare environment
EMS Compliance Table (Table 2-5)
Table 2 - Enclosure Port
Phenomenon
Basic EMC standard
Immunity test levels
Home healthcare environment
Electrostatic
Discharge
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Radiated RF EM field
IEC 61000-4-3
20 V/m
26 MHz-2.5 GHz
80% AM at 1 kHz
10 V/m
80 MHz-2.7 GHz
80% AM at 1 kHz
Proximity fields from RF
wireless communications
equipment
IEC 61000-4-3
Refer to table 3
Rated power frequency
magnetic fields
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz