Figure 8: Setting the External Adjustment Rings
8.
Begin with the upper external adjustment ring at the top of the external adjustment body (
).
Welker recommends using a felt tip pen to mark the front of both external adjustment rings so that it is clear when the
adjustment rings have made a full rotation.
9.
Using the provided external adjustment tool, turn the upper adjustment ring to the desired volume (
One full rotation of the upper external adjustment ring is approximately 1 cc if the inFlow™ ACE is equipped with the D-Style
collection head. Each hole in the upper external adjustment ring is equal to approximately 0.10 cc.
10.
Turn the lower external adjustment ring until it meets the upper external adjustment ring. Tighten firmly to prevent
slippage that might lead to inconsistencies in the sample volume.
The lower external adjustment ring and O-ring are used to lock the upper external adjustment ring in place to prevent slippage
and subsequent changes in the sample volume during sampling.
11.
Turn ON and actuate the hydraulic or pneumatic supply to collect a sample from the sample outlet to ensure that the
sample volume collected is the same as the desired sample volume.
Welker recommends the Welker Checkpoint™ Sample Bite Verification Panel for use with this unit.
12.
As necessary, repeat steps 6–11 to further adjust the sample volume.