4-46
4.12.2 How to Store/Delete Sweep References
The Sweep References soft key allows the user to select, store or
delete sweep references. After a sweep measurement is made, a
sweep reference can be stored. The unit must remain in the Sweep
screen for a minimum amount of time before a reference can be
stored. When first entering the Sweep screen, a wait indicator
appears above the sweep graph. A reference cannot be stored until
the wait indicator goes away. To store a Sweep Reference, press the
File
key followed by the
Sweep References
soft key. Next, press the
Store
soft key to name and store the sweep measurement as a
reference.
NOTE: The Store soft key is only active if Sweep was the most
recent mode selected prior to pressing the File key.
Sweep references can be deleted, by pressing the
Delete
soft key.
To select a reference, use the
UP
and
DOWN
arrow keys to highlight
the desired reference and then press the
ENTER
or
SWEEP
key to
return to the Sweep screen.
If the channel plan has changed since a reference has been stored,
the reference is obsolete and can no longer be used. If an obsolete
reference is selected, a warning message will appear upon entering