4-10
C/N Calibration
To increase the accuracy of the C/N measurements, a noise floor
calibration is performed. This allows the user to characterize the
noise floor of the unit. To perform the calibration, highlight the
selection and press the
ENTER
key. The unit will provide a prompt
to ensure that no cable is connected to the RF input. Once verified,
press the
OK
soft key to perform the calibration. The unit will
display the system noise level when the calibration is complete.
4.3.3 How To Setup Your Channel Plan
The procedure explained in Section 4.2 should be performed when
first using the Model 3SR Receiver.