IS-HF-M100+-29293  Rev 5

 

10

 

Operation

Home Screen

All operations start on the Home screen. Refer to the above figure for 

all operations.

•  When the instrument is in Automatic Readings, Update Reading 

button will not show. This is only used for manual updates.

•  The current reading is always shown in the large number format.
•  Toggle between NTU and FNU by touching the button.
•  Pressing Sample ID brings up a key pad to allow you to provide 

a name.

•  Sample ID shows up to five previous readings. 100 readings 

per Sample ID are saved. Once logging is complete, the 

Home

 

screen will say "100 points captured." See 

Trend

 screen below 

for more information.

•  The instrument can be turned off manually even if Auto Power 

Down is selected.

Trend Screen

•  Select Sample ID first.
•  After a reading session in the Home screen, if Trend screen is 

selected, the Reading List from the Home screen is saved under 

the Sample ID in the Trend screen and cleared from the Home 

screen. If a new session is started in the Home screen under the 

same Sample ID, the readings will be appended to the session 

readings.

•  If a USB memory stick is connected, Session Readings shown will 

be transferred to the memory stick after pressing 

Download/Print

.

•  If the USB printer is connected, Sessions Readings shown will be 

printed after pressing 

Download/Print

.

•  Sample ID will be saved until deleted. See Delete Samples 

section for more information.

Update Reading

Current Reading

Time

Sample ID 

(Name)

NTU/FNU

Diagnostic 

Banner

On/Off

Reading List

Setup

Calibration

Trend

Home

Data Retrieval

•  Data logging can only save 100 points per Sample ID.
•  All data is saved in CSV format and is date and time stamped.
•  Sample ID Data can be downloaded to a USB memory stick or 

printed in the 

Trend

 screen after pressing 

Download/Print

.

•  Calibration data can be downloaded to a USB memory stick in 

the Calibration Download menu by pressing the 

 button. See 

Calibration Download section for more information.

•  Logged data can be downloaded to a USB memory stick in the 

Log Download menu by pressing the 

 button. 

USB Connection

The M100+ has two USB connections: a USB-A on the side and a 

USB-B on the back. Only the side connection is active at this time. The 

USB-B on the back of the instrument may be used in a future firmware 

update.
The USB-A connection can be used to upload new software or 

download logged data and calibration reports via a USB memory stick.
Firmware updates can be made by loading the firmware onto a USB 

memory stick and inserting it into the USB-A connector. The update 

will not affect configurations or stored data, however it is always 

advisable to check the configuration after an update as menus may 

have changed.
Be sure to check our website, www.hfscientific.com, for software 

updates.

Screen 

Selection

Sample ID 

Selection

To get session 

readings

Time

Download/

Print

Graph

To keep the USB contacts from getting contaminated, a USB 

plug cover is provided. It is not recommended that this cover is 

replaced whenever the USB is not being used. The USB-B plug 

on the back is not active at this time and the USB plug cover 

should remain in place to prevent damage.

NOTICE

Содержание HF scientific M100 28060

Страница 1: ...een Calibration 6 Set Date Time 6 Delete Samples 6 Data Resolution 6 Offset 7 Device ID 7 Calibration Reminder 7 Calibration Download 7 Firmware Update 7 Reset Defaults 7 About Device 7 Calibrating th...

Страница 2: ...ould not permit anyone to touch this equipment unless they are over 18 years of age are adequately trained and supervised and have read and understood this manual Owners should also ensure that no una...

Страница 3: ...t in death or serious injury This symbol identifies hazards which if not avoided could result in death or serious injury This symbol identifies hazards which if not avoided could result in minor or mo...

Страница 4: ...ns for the M100 are valid only for measurement of static non flowing samples Touch Screen Standard Inserted in Optics USB Connector Unpacking and Inspection of the Instrument Remove all items from pac...

Страница 5: ...gned to provide the ability to customize the instrument according to your needs at any time during normal operation The Setup screen has been split into sub menus to facilitate instrument configuratio...

Страница 6: ...ation The screen has been factory adjusted but can be adjusted if the alignment is off The screen may need to be calibrated if new software has been downloaded It is a 5 position setup 1 Press the Set...

Страница 7: ...sert the USB memory stick that contains the software into the USB port 3 If the firmware revision on the memory stick is older than the firmware currently installed the instrument will ask if you wish...

Страница 8: ...ded as general guidelines for proper care of ProCal standards Standards should only be handled by the top black cap or by the very bottom of the cuvette Keep cuvette glass clean of fingerprints and de...

Страница 9: ...n on the left side of the screen 2 Insert the 4000 NTU standard 3 Press the Command button in the middle of the screen to start calibration A countdown from 30 will appear on the screen When the count...

Страница 10: ...stic Banner On Off Reading List Setup Calibration Trend Home Data Retrieval Data logging can only save 100 points per Sample ID All data is saved in CSV format and is date and time stamped Sample ID D...

Страница 11: ...Email HF Info wattswater com www hfscientific com Routine Maintenance Routine maintenance should include calibration of the instrument See Calibration section in this manual for more information Clea...

Страница 12: ...al or consequential damages including without limitation lost profits or the cost of repairing or replacing other property which is damaged if this product does not work properly other costs resulting...

Отзывы: