manualshive.com logo in svg
background image

IS-HF-M100+-29293  Rev 5

 

9

 

Calibration Procedures (continued)

1.  Select the calibration function by pressing the 

Calibrate

 icon.

2.  Select the appropriate calibration standard (Formazin or ProCal) and 

then press 

 to save the selection.

3.  The calibration menu will be displayed as shown above. On the 

left-hand side are selections for the range for calibration. Select the 

highest value required for your application.

4.  Ensure standards are indexed as described in Indexing Calibration 

Cuvettes section prior to calibration.

5.  Press the command button. The command button dialogue will 

guide you through the calibration.

Calibrating using ProCal Standards

1.  Press the High Range A button on the left side of the screen.
2.  Insert the High Range A standard.
3.  Press the Command button in the middle of the screen to start 

calibration. A countdown from 30 will appear on the screen. When 

the countdown is complete, the screen will request the High Range 

B.

4.  Press the High Range B button on the left side of the screen.
5.  Insert the High Range B standard. 
6.  Press the Command button to continue the calibration. A 

countdown from 30 will appear on the screen. When the 

countdown is complete, the screen will request the 1000 NTU 

standard.

7.  Press the 1000 NTU button on the left side of the screen.
8.  Insert the 1000 NTU standard.
9.  Press the Command button to continue the calibration. A 

countdown from 60 will appear on the screen. When the 

countdown is complete, the screen will request the 10 NTU 

standard.

10. Press the 10 NTU button on the left side of the screen.

Select 

Range

Command 

button: 

Follow 

directions 

that appear 

here for 

calibration.

Formazin Calibration Screen

ProCal Calibration Screen

11. Insert the 10 NTU standard.
12. Press the Command button to continue the calibration. A 

countdown from 60 will appear on the screen. When the 

countdown is complete, the screen will request the 0.02 NTU 

standard.

13. Press the 0.02 NTU button on the left side of the screen. 
14. Inert the 0.02 NTU standard.
15. Press the Command button to continue the calibration. A 

countdown from 30 will appear on the screen. 

When the countdown is complete, all buttons on the left side of 

the screen should be green and the Command button should say 

Calibration Good.

Calibrating Using Formazin Standards

1.  Press the 4000 button on the left side of the screen.
2.  Insert the 4000 NTU standard.
3.  Press the Command button in the middle of the screen to start 

calibration. A countdown from 30 will appear on the screen. When 

the countdown is complete, the screen will request the 1000 NTU 

standard.

4.  Press the 1000 NTU button on the left side of the screen. 
5.  Insert the 1000 NTU standard. 
6.  Press the Command button to continue the calibration. A 

countdown from 60 will appear on the screen. When the 

countdown is complete, the screen will request the 10 NTU 

standard.

7.  Press the 10 NTU button on the left side of the screen.
8.  Insert the 10 NTU standard.
9.  Press the Command button to continue the calibration. A 

countdown from 60 will appear on the screen. When the 

countdown is complete, the screen will request the 0.02 NTU 

standard.

10. Press the 0.02 NTU button on the left side of the screen.
11. Insert the 0.02 NTU standard.
12. Press the Command button to continue the calibration. A 

countdown from 30 will appear on the screen. 

When the countdown is complete, all buttons on the left side of 

the screen should be green and the Command button should say 

Calibration Good.

Alternate Calibration Ranges

Alternately you could choose to calibrate to 1000 NTU, which requires 

1000 NTU, 10 NTU and 0.02 NTU. Press the 1000 button on the left 

side and start at step 5 of the previous section.
A calibration for only 10 NTU requires only the 10 NTU and 0.02 NTU 

standards. Press the 10 NTU button on the left side and start at step 8 

of the previous section.
For the two alternate ranges, these will still read the full range with a 

reduced accuracy above the range of calibration.

Failed Calibration

In the case of a failed calibration, the last good calibration is restored 

once the user exits to the Home screen. A failure will be evident when 

the calibration screen shows all red.
The instrument can be used with this calibration with a potentially 

reduced accuracy. The bad calibration will be noted in the calibration 

log and as a reminder, a yellow diagnostic banner will show on the 

home screen. The yellow banner indicates that the last calibration had 

failed and that the instrument is using the previous calibration. The only 

way to remove this indication is to successfully complete a calibration.
Check the standards for cleanliness. If everything looks good, attempt 

the calibration again. If the calibration fails after the second attempt, 

the most likely cause is that the standards are bad or the lamp needs 

replacement.

Содержание HF scientific M100 28060

Страница 1: ...een Calibration 6 Set Date Time 6 Delete Samples 6 Data Resolution 6 Offset 7 Device ID 7 Calibration Reminder 7 Calibration Download 7 Firmware Update 7 Reset Defaults 7 About Device 7 Calibrating th...

Страница 2: ...ould not permit anyone to touch this equipment unless they are over 18 years of age are adequately trained and supervised and have read and understood this manual Owners should also ensure that no una...

Страница 3: ...t in death or serious injury This symbol identifies hazards which if not avoided could result in death or serious injury This symbol identifies hazards which if not avoided could result in minor or mo...

Страница 4: ...ns for the M100 are valid only for measurement of static non flowing samples Touch Screen Standard Inserted in Optics USB Connector Unpacking and Inspection of the Instrument Remove all items from pac...

Страница 5: ...gned to provide the ability to customize the instrument according to your needs at any time during normal operation The Setup screen has been split into sub menus to facilitate instrument configuratio...

Страница 6: ...ation The screen has been factory adjusted but can be adjusted if the alignment is off The screen may need to be calibrated if new software has been downloaded It is a 5 position setup 1 Press the Set...

Страница 7: ...sert the USB memory stick that contains the software into the USB port 3 If the firmware revision on the memory stick is older than the firmware currently installed the instrument will ask if you wish...

Страница 8: ...ded as general guidelines for proper care of ProCal standards Standards should only be handled by the top black cap or by the very bottom of the cuvette Keep cuvette glass clean of fingerprints and de...

Страница 9: ...n on the left side of the screen 2 Insert the 4000 NTU standard 3 Press the Command button in the middle of the screen to start calibration A countdown from 30 will appear on the screen When the count...

Страница 10: ...stic Banner On Off Reading List Setup Calibration Trend Home Data Retrieval Data logging can only save 100 points per Sample ID All data is saved in CSV format and is date and time stamped Sample ID D...

Страница 11: ...Email HF Info wattswater com www hfscientific com Routine Maintenance Routine maintenance should include calibration of the instrument See Calibration section in this manual for more information Clea...

Страница 12: ...al or consequential damages including without limitation lost profits or the cost of repairing or replacing other property which is damaged if this product does not work properly other costs resulting...

Отзывы: