Troubleshooting

M100+ Fault Detection

The M100+ performs continuous diagnostic monitoring. The diagnostic 

indicator will display any errors or warnings on the Home screen.

•  A green indicator shows that everything is good.
•  A red indicator denotes an error.

- Potential errors are a bad lamp or the reading is taking place 

in a band that has not been calibrated.

•  A yellow indicator is a warning and could show either a faulty 

database or the device plugged into the USB is incompatible.

Diagnostic Table

Symptom

Cause

Solution

Readings are higher than 

expected.

Bubbles in solution.

Clean the inside and outside 

of the cuvette with a detergent 

solution. Rinse with deionized 

water. Do not shake samples.

Instrument out of 

calibration.

Recalibrate. Refer to Calibration 

section in this manual.

USB download not working.

USB memory stick is not 

fully inserted.

Insert memory stick fully.

USB Memory stick is not 

formatted.

Format memory stick to FAT32.

Readings are erratic.

Bubbles in solution.

Clean the inside and outside 

of the cuvette with a detergent 

solution. Rinse with deionized 

water. Do not shake samples.

Readings are lower than 

expected.

Instrument out of 

calibration.

Recalibrate. Refer to Calibration 

section in this manual.

Screen calibration not 

working.

Incorrect screen 

calibration stored.

Hold upper (time) banner on any 

screen for 5 seconds to load 

the default screen calibration. 

Then go to screen calibration 

and follow instructions in Touch 

Screen Calibration section in this 

manual.

Technical and Customer Assistance

If you need assistance with this instrument or to order replacement 

parts, please contact HF scientific Customer Service or Technical 

Service Departments.

HF scientific 

16260 Airport Park Drive, Suite 140 

Fort Myers, Florida 33913 

Phone: (239) 337-2116 

Toll Free: (888) 203-7248 

Fax: (239) 454-0694 

Email: [email protected] 

www.hfscientific.com

Routine Maintenance

Routine maintenance should include calibration of the instrument. See 

Calibration section in this manual for more information.

Cleaning the Cuvette

Measurement cuvettes used for the sample should be clean and free of 

marks or scratches.
Clean the interior and exterior of the cuvettes with a detergent solution 

and then rinse several times with distilled or deionized water. Ensure 

the cuvettes are clean prior to use.
If cuvettes become scratched or stained, they should be replaced. 

Replacement cuvettes are available from HF scientific. Order catalog 

#50051 (3 pack of cuvettes with caps) or 50052 (10 pack of cuvettes 

with caps).

Replacing the Source Lamp

The infrared and white light source lamps in the M100+ instruments 

are designed for long life. However, these lamps are not covered by 

the warranty. If the lamp should need replacement, please contact HF 

scientific's Service Department for assistance. The lamps can also be 

purchased with instructions to change. Calibration will be required if a 

lamp change is made.

Battery

This instrument uses a battery for backup of the clock feature with an 

expected life span of over 3 years. The battery is non-rechargeable 

and no attempt to charge it should be made. If the battery needs to be 

replaced, contact HF scientific to arrange for service.

Accessories and Replacement Parts

Catalog #

Description

Photo

24082S

Replacement Lamp Assembly, White Light w/ 

Instructions

21396S

Replacement Lamp Assembly, IR Light w/ 

Instructions

29293

Operating Manual, M100+

n/a

39957

ProCal Calibration Kit, 0.02, 10 & 1000 NTU

39940

ProCal Calibration Kit, 0.02, 10, 1000, High 

Range A & High Range B (W/L)

39941

ProCal Calibration Kit, 0.02, 10, 1000, High 

Range A & High Range B, (IR)

39825

ProCal Calibration Standard, 10 NTU, 125ml

50040

Formazin Stock Kit

n/a

70914

Formazin Stock Solution, 4000 NTU, 500 ml

50051

Replacement Cuvette - 3 Pack

50052

Replacement Cuvette - 10 Pack

29297S

US Wall Mount Power Supply

n/a

29301S

EU Wall Mount Power Supply

n/a

Содержание HF scientific M100 28060

Страница 1: ...een Calibration 6 Set Date Time 6 Delete Samples 6 Data Resolution 6 Offset 7 Device ID 7 Calibration Reminder 7 Calibration Download 7 Firmware Update 7 Reset Defaults 7 About Device 7 Calibrating th...

Страница 2: ...ould not permit anyone to touch this equipment unless they are over 18 years of age are adequately trained and supervised and have read and understood this manual Owners should also ensure that no una...

Страница 3: ...t in death or serious injury This symbol identifies hazards which if not avoided could result in death or serious injury This symbol identifies hazards which if not avoided could result in minor or mo...

Страница 4: ...ns for the M100 are valid only for measurement of static non flowing samples Touch Screen Standard Inserted in Optics USB Connector Unpacking and Inspection of the Instrument Remove all items from pac...

Страница 5: ...gned to provide the ability to customize the instrument according to your needs at any time during normal operation The Setup screen has been split into sub menus to facilitate instrument configuratio...

Страница 6: ...ation The screen has been factory adjusted but can be adjusted if the alignment is off The screen may need to be calibrated if new software has been downloaded It is a 5 position setup 1 Press the Set...

Страница 7: ...sert the USB memory stick that contains the software into the USB port 3 If the firmware revision on the memory stick is older than the firmware currently installed the instrument will ask if you wish...

Страница 8: ...ded as general guidelines for proper care of ProCal standards Standards should only be handled by the top black cap or by the very bottom of the cuvette Keep cuvette glass clean of fingerprints and de...

Страница 9: ...n on the left side of the screen 2 Insert the 4000 NTU standard 3 Press the Command button in the middle of the screen to start calibration A countdown from 30 will appear on the screen When the count...

Страница 10: ...stic Banner On Off Reading List Setup Calibration Trend Home Data Retrieval Data logging can only save 100 points per Sample ID All data is saved in CSV format and is date and time stamped Sample ID D...

Страница 11: ...Email HF Info wattswater com www hfscientific com Routine Maintenance Routine maintenance should include calibration of the instrument See Calibration section in this manual for more information Clea...

Страница 12: ...al or consequential damages including without limitation lost profits or the cost of repairing or replacing other property which is damaged if this product does not work properly other costs resulting...

Отзывы: