14.1
Pump external interface parameters
Parameter
Limits
Units
Comment
Sym Min Nom Max
Digital input voltage high
VD
IH
5
24
V
Leak, Stop, Pressure,
Frequency
Digital input voltage low
VD
IL
0
0.8
V
Leak, Stop, Pressure,
Frequency
Digital input voltage absolute
maximum
VD
in
-30
30
V
Non operational
Digital input resistance
RD
in
10
110
kΩ
110K for ≤ 5V
Frequency range
F
max
1
1000
Hz
Flow sensor use
Analogue input, voltage mode
VA
in
-15
10
30
V
0-10V range (100R source
impedance)
Analogue input, voltage mode
RVA
in
34.4
kΩ
±3%
Analogue input measurement
range
I
in
0
25
mA
Analogue input current absolute
maximum
IA
in
-50
28
mA
Dissipation limit
Analogue input voltage absolute
maximum
VA
in
0
7.0
V
Dissipation limit
Analogue input resistance
RI
IN
250
270
Ω
250R Sense Res.
Analogue input filter bandwidth
BW
67
Hz
-6dB Bandwidth
22V supply output
V
aux
18
30
V
Un-regulated
24V isolated supply output
V24
24
22V/24V supply load current
80
mA
Self-resetting fuse
34
m-630bpn-en-08
Содержание 630 Bp
Страница 21: ...1 2 m 630bpn en 08 21...
Страница 46: ...14 7 Network topology Star network Ring network 46 m 630bpn en 08...
Страница 105: ...22 5 Flow Sensor Reading The flow sensor value can be read via the Flow sensor reading screen m 630bpn en 08 105...
Страница 107: ...Press SELECT to choose an option Press EXIT to return to the screen from where the MENU was called m 630bpn en 08 107...
Страница 130: ...24 Help 24 1 Help Select Help from the main menu to access the help screens 130 m 630bpn en 08...
Страница 150: ...Number Spare Description 8 TT0005 10mm 3 8in spanner 9 MRA3026A Track assembly 150 m 630bpn en 08...