General guide to cleaning with solvents
Chemical
Cleaning precautions
Aliphatic hydrocarbons
Remove guard. Minimize rotor cap and clutch boot
exposure to less than one minute (risk of attack).
Aromatic hydrocarbons
Remove guard. Minimize rotor cap and clutch boot
exposure to less than one minute (risk of attack).
Ketone solvents
Remove guard. Minimize rotor cap and clutch boot
exposure to less than one minute (risk of attack).
Halogenated/chlorinated solvents
Not recommended: possible risk to polycarbonate
tube clamp adjusters and polypropylene tube clamp
locators.
Alcohols, general
No precaution necessary.
Glycols
Minimize rotor cap and clutch boot exposure to less
than one minute (risk of attack).
Estersolvents
Remove guard. Minimize rotor cap and tube clamp
location cap exposure to less than one minute (risk of
attack).
Ether solvents
Not recommended: possible risk to polycarbonate
tube clamp adjusters and polypropylene tube clamp
locators.