ReNu 20, ReNu 60 or ReNu 120
IMPORTANT: Set the pressure valve to the ‘in-use’ position before
installing the pumphead.
In ‘Transportation position’ Leak detector will not function with
discharge pressure below 1 bar (15 psi).
Note:
In this manual, we have shown removal and replacement of the left-hand mounted
pumphead. Replacing the right-hand mounted pumphead is the identical procedure.
Removing the pumphead
1.
Drain down.
2.
Ensure that there is no pressure in the pipeline.
3.
Isolate the pump from the mains power supply.
4.
Ensure that protective clothing and eye protection are worn if hazardous products
have been pumped.
5.
Remove input and output connections from the pumphead (protecting the pump
from any process fluid spillages).
6.
Release the pumphead retaining lever.