3. Grasp the leak sensor by its serrations and pull it upward to remove it from its reservoir.
Figure 3–24: Leak sensor serrations
1
Serrations
4. Unpack the new leak sensor.
5. Align the leak sensor’s T-bar with the slot in the side of the leak sensor reservoir and slide
the leak sensor into place.
Figure 3–25: Aligning T-bar with slot
T-bar
Slot in leak sensor reservoir
6. Connect the leak sensor connector to the front of the device.
Figure 3–26: Attaching leak sensor connector
Leak sensor connector
Leak sensor port on front of device
7. Close the instrument door.
August 20, 2021, 715005003 Ver. 04 (Previously released as Revision C)