Figure 2–2: UNIFI health system - RDa
Link to Online help
Issues and Information tabs
Show Information button
Health card
Issues tab
The following are displayed under the
Issues
tab:
• Activities - The instrument is in a
Getting Ready
state. Activities are actions the instrument
is performing to get to a
Ready
state.
• Warnings - The instrument is in a
Check
state. Warnings are issues observed by the health
system that you can rectify through user maintenance activities.
• Errors - The instrument is in an
Error
state and requires the help of a Waters service
engineer. The panel on the instrument displays the words
Call Service
.
Information tab
Additional information is displayed on the Summary page under the
Information
tab. These
alerts do not affect the instrument health or the operation of the instrument; instead, they can
help when multiple scientists are using the system. For example, you can see if the Source
Pressure Test was overridden or if calibrations were not run for specific slots.
August 20, 2021, 715005003 Ver. 04 (Previously released as Revision C)
Page 24