16
Diagnostics and Calibration
2
Status light blinks and
lamp light on
Failed startup
diagnostics
Reseat and check alignment of flow
cell.
Flush the flow cell (Section 3.1.1,
Flushing the Flow Cell).
Insufficient energy
reaching photodiode
array because of air
bubble, or dirty flow
cell can cause shutter
diagnostic to fail
Flush the flow cell (Section 3.1.1,
Flushing the Flow Cell).
To prevent air bubbles from forming,
check that there is a 1- to 2-foot (30- to
60-cm) length of 0.009-inch
(0.23-mm) ID tubing connected to the
detector waste outlet.
Weak lamp
Replace lamp (Section 3.1.2,
Removing the Flow Cell).
Shutter failure
message
Shutter failed
Run the Shutter diagnostic. For details,
see Millennium
32
Help.
Detector not
responding to
Millennium
32
Workstation
Detector not
connected to
busLAC/E or to
LAC/E
32
Acquisition
Server in the
Millennium
32
Workstation
Check IEEE-488 cable connections,
tighten connectors.
Incorrect IEEE-488
address
1. Ensure that the 2996 Detector
IEEE-488 address is unique and
within the range 2 to 29 (see the
Millennium
32
System Installation
and Configuration Guide).
2. Rescan the IEEE-488 bus. For
details, see Millennium
32
Help.
Table 2-1 2996 Detector Troubleshooting
(Continued)
Symptom
Possible Cause
Corrective Action
Содержание 2996
Страница 1: ...Waters 2996 PDA Detector Operator s Guide 34 Maple Street Milford MA 01757 71500023202 Revision C...
Страница 14: ...xiv Table of Contents...
Страница 16: ...xvi List of Figures...
Страница 18: ...xviii List of Tables...
Страница 22: ...xxii...
Страница 36: ...14 Installation 1...
Страница 62: ...40 Principles of the 2996 PDA Detector Optics 4...
Страница 72: ...50 Spectral Contrast Theory 5...
Страница 74: ...52 Detector Specifications A...
Страница 76: ...54 Spare Parts B...
Страница 84: ...62 Index I N D E X...