Mobile Phase Absorbance
55
C
Appendix C
Mobile Phase Absorbance
This appendix lists the absorbances at several wavelengths for commonly used mobile
phases. Choose your mobile phase carefully to reduce baseline noise.
The best mobile phase for your application is one that is transparent at the chosen
detection wavelengths. Such a mobile phase ensures that any absorbance is due only to the
sample. Absorbance by the mobile phase also reduces the linear dynamic range of the
detector by the amount of absorbance that is autozeroed out. Wavelength, pH, and
concentration of the mobile phase will affect its absorbance. Examples of several mobile
phases are provided in Table C-1.
Table C-1 Mobile Phase Absorbance Measured Against Air or Water
Absorbance at Specified Wavelength (nm)
200
205
210
215
220
230
240
250
260
280
Solvents
Acetonitrile
0.05
0.03
0.02
0.01
0.01
<0.01
—
—
—
—
Methanol (not
degassed)
2.06
1.00
0.53
0.37
0.24
0.11
0.05
0.02
<0.01
—
Methanol
(degassed)
1.91
0.76
0.35
0.21
0.15
0.06
0.02
<0.01
—
—
Isopropanol
1.80
0.68
0.34
0.24
0.19
0.08
0.04
0.03
0.02
0.02
Unstablized
Tetrahydrofuran
(THF, fresh)
2.44
2.57
2.31
1.80
1.54
0.94
0.42
0.21
0.09
0.05
Unstablized
Tetrahydrofuran
(THF, old)
>2.5
>2.5
>2.5
>2.5
>2.5
>2.5
>2.5
>2.5
2.5
1.45
Содержание 2996
Страница 1: ...Waters 2996 PDA Detector Operator s Guide 34 Maple Street Milford MA 01757 71500023202 Revision C...
Страница 14: ...xiv Table of Contents...
Страница 16: ...xvi List of Figures...
Страница 18: ...xviii List of Tables...
Страница 22: ...xxii...
Страница 36: ...14 Installation 1...
Страница 62: ...40 Principles of the 2996 PDA Detector Optics 4...
Страница 72: ...50 Spectral Contrast Theory 5...
Страница 74: ...52 Detector Specifications A...
Страница 76: ...54 Spare Parts B...
Страница 84: ...62 Index I N D E X...