Diagnostics
15
2
Chapter 2
Diagnostics and Calibration
The Waters 2996 Photodiode Array Detector automatically runs a series of internal
diagnostics upon start up. The indicator LEDs on the front of the detector and messages at
the Millennium
32
Workstation show the results of the start up internal diagnostics
(Figure 1-8).
If you need to determine the cause of a problem during operation of the detector, you can
run the same internal diagnostics from the Millennium
32
Workstation. Additional
information about the performance of the detector is also available through the PDA
Calibration window, accessed from Run Samples in the Millennium
32
software.
If you encounter a problem that you cannot troubleshoot (Section 2.1, Diagnostics),
contact Waters Technical Service at (800) 252-4752, U.S. and Canadian customers only.
Other customers, call your local Waters subsidiary or your local Waters Technical Service
Representative in Milford, Massachusetts (U.S.A.) for assistance.
2.1 Diagnostics
Refer to Table 2-1 to troubleshoot problems encountered during startup diagnostics and
during detector operation.
Table 2-1 2996 Detector Troubleshooting
Symptom
Possible Cause
Corrective Action
Both LEDs off
No power
1. Check line cord connections.
2. Check outlet for power.
Blown fuse
Replace fuse (Section 3.3, Replacing
the Fuses).
Status light blinks and
lamp light is off
The 2996 is running
Confidence tests.
Содержание 2996
Страница 1: ...Waters 2996 PDA Detector Operator s Guide 34 Maple Street Milford MA 01757 71500023202 Revision C...
Страница 14: ...xiv Table of Contents...
Страница 16: ...xvi List of Figures...
Страница 18: ...xviii List of Tables...
Страница 22: ...xxii...
Страница 36: ...14 Installation 1...
Страница 62: ...40 Principles of the 2996 PDA Detector Optics 4...
Страница 72: ...50 Spectral Contrast Theory 5...
Страница 74: ...52 Detector Specifications A...
Страница 76: ...54 Spare Parts B...
Страница 84: ...62 Index I N D E X...