Fluidic Connections
9
1
Electrical Specifications
Before you connect an external device to an event input or output terminal, refer to the
electrical specifications in Table 1-2 and Table 1-3.
1.5 Fluidic Connections
Required Materials
• 5/16-inch open-end wrench
• 0.009-inch (0.23 mm) ID stainless steel tubing (included in the 2996 Detector
Startup Kit)
Table 1-2 Event In Terminal Specifications on TTL or Switch Closure
Parameter
Specification
Low trigger
<1.8 V
High trigger
>3.0 V
Protected to
±30 V
Minimum pulse width
30 msec (this may limit compatibility
with valves that require a quick pulse)
Maximum current
5 mA
Table 1-3 Event Out Terminal Specifications on Contact Closure
Parameter
Specification
Maximum power
10 W
Maximum current
0.5 A at 20 V
Maximum voltage
24 V RMS
STOP
Attention:
To avoid damage to the 2996 Detector electronics, be sure you make the
proper electrical connections as outlined in this section.
Caution:
To avoid chemical hazards, always observe safe laboratory practices when
handling solvents. Refer to the Material Safety Data Sheets for the solvents in use.
Содержание 2996
Страница 1: ...Waters 2996 PDA Detector Operator s Guide 34 Maple Street Milford MA 01757 71500023202 Revision C...
Страница 14: ...xiv Table of Contents...
Страница 16: ...xvi List of Figures...
Страница 18: ...xviii List of Tables...
Страница 22: ...xxii...
Страница 36: ...14 Installation 1...
Страница 62: ...40 Principles of the 2996 PDA Detector Optics 4...
Страница 72: ...50 Spectral Contrast Theory 5...
Страница 74: ...52 Detector Specifications A...
Страница 76: ...54 Spare Parts B...
Страница 84: ...62 Index I N D E X...