Installation Site Requirements
1
1
Chapter 1
Installation
The Waters
®
2996 Photodiode Array (PDA) Detector operates in any standard laboratory
environment. The detector requires electrical power, sample and waste fluidic lines, and
either the Millennium®
32
or MassLynx™ workstations. Optional connections on the
detector rear panel allow communication with chart recorders, data integrators, and other
instruments that are not compatible with Millennium
32
or MassLynx software control.
1.1 Installation Site Requirements
Install the 2996 PDA Detector at a site that meets the specifications in Table 1-1 and
Figure 1-1.
Table 1-1 Site Requirements
Factor
Specification
Ambient temperature
4 to 40 °C (39 to 104 °F), not to vary more
than 1 °C per hour (to prevent drift)
Relative humidity
20 to 80%, noncondensing
Bench space
Width: 11.5 in. (29 cm)
Depth: 24 in. (61 cm)
Height: 8.5 in. (22 cm)
Bench support
Capable of supporting 31.5 pounds (14.3 kg)
Clearance
At least 4 in. (10 cm) on the back side for
ventilation
Power
Grounded ac, 100 to 240 Vac
Содержание 2996
Страница 1: ...Waters 2996 PDA Detector Operator s Guide 34 Maple Street Milford MA 01757 71500023202 Revision C...
Страница 14: ...xiv Table of Contents...
Страница 16: ...xvi List of Figures...
Страница 18: ...xviii List of Tables...
Страница 22: ...xxii...
Страница 36: ...14 Installation 1...
Страница 62: ...40 Principles of the 2996 PDA Detector Optics 4...
Страница 72: ...50 Spectral Contrast Theory 5...
Страница 74: ...52 Detector Specifications A...
Страница 76: ...54 Spare Parts B...
Страница 84: ...62 Index I N D E X...