Making Fluidic Connections
37
2
Figure 2-2 IEEE-488 Connection Example
2.
Connect the dual-receptacle end of the 2-meter IEEE-488 cable to the IEEE-488
port on one of the IEEE-488 devices (pump, column heater, autosampler, or
detector).
3.
Connect one end of a 1-meter IEEE-488 cable (with dual-receptacle connectors at
each end) to the cable receptacle on the first device. Connect the other end of the
cable to the IEEE-488 port on the next device.
4.
Repeat step 3 for each additional IEEE-488 device.
5.
Make sure that all IEEE-488 cable connector screws are finger tight.
2.4 Making Fluidic Connections
This section describes how to make fluidic connections to a Waters 1525 Pump by:
•
Making eluent supply connections
•
Installing different mixer configurations
•
Installing an optional manual injector
•
Installing an optional column heater
•
Making outlet connections
•
Connecting fluid waste lines
2.4.1 Making Eluent Supply Connections
Follow the instructions in this section to connect a 1525µ or 1525EF Pump inlet to an
eluent reservoir.
GPIB Card in
Workstation
Pump
Detector
2-Meter
IEEE-488
Cable
1-Meter
IEEE-488
Cables
IEEE-488
Connectors
Column
Heater
Autosampler