Waters TQ Detector IVD

Operator’s Guide 

71500126802IVD Revision C

Copyright © Waters Corporation 2006–2012

All rights reserved

Содержание TQ

Страница 1: ...Waters TQ Detector IVD Operator s Guide 71500126802IVD Revision C Copyright Waters Corporation 2006 2012 All rights reserved...

Страница 2: ...ctions INSIGHT and Waters are registered trademarks of Waters Corporation ACQUITY IntelliStart IonSABRE MassLynx T Wave UPLC and THE SCIENCE OF WHAT S POSSIBLE are trademarks of Waters Corporation Nal...

Страница 3: ...tact information Contacting medium Information Internet The Waters Web site includes contact information for Waters locations worldwide Visit www waters com Telephone and fax From the USA or Canada ph...

Страница 4: ...bar 100 psi during analyses that require flammable solvents Connect to the LC output with a gas fail connector to stop the LC solvent if the nitrogen supply fails Warning To confirm the integrity of t...

Страница 5: ...ector high temperature hazard Warning To avoid burn injuries avoid touching the source enclosure with your hand when operating or servicing the instrument NEBULIZER NEBULIZER APPI APPI PROBE PROBE HV...

Страница 6: ...003 Cet appareil num rique de la classe A est conforme la norme NMB 003 Electrical power safety notice Do not position the instrument so that it is difficult to operate the disconnecting device Safet...

Страница 7: ...ipment misuse notice If the equipment is used in a manner not specified by the manufacturer the protection provided by the equipment may be impaired Safety advisories Consult Appendix A for a comprehe...

Страница 8: ...plies with all applicable European Community directives Australia C Tick EMC Compliant Confirms that a manufactured product complies with all applicable United States and Canadian safety requirements...

Страница 9: ...calibration methods using at least five standards to generate a standard curve The concentration range for standards must cover the entire range of quality control samples typical specimens and atypic...

Страница 10: ...When analyzing samples from a complex matrix such as serum plasma whole blood and so on note that the matrix components can adversely affect LC MS results enhancing or suppressing ionization To minim...

Страница 11: ...in commercial that is nonresidential locations and can be directly connected to a low voltage power supply network In a domestic environment the equipment may cause radio interference You may need to...

Страница 12: ...12 71500126802IVD Rev C EC authorized representative Waters Corporation Micromass UK Ltd Floats Road Wythenshawe Manchester M23 9LZ United Kingdom Telephone 44 161 946 2400 Fax 44 161 946 2480 Contact...

Страница 13: ...er safety notice vi Safety hazard symbol notice vi Equipment misuse notice vii Safety advisories vii Operating this instrument viii Applicable symbols viii Intended use ix Calibrating ix Quality contr...

Страница 14: ...5 Starting the instrument 36 Configuring IntelliStart 38 Verifying the instrument s state of readiness 38 Tuning and calibration information 38 Running the instrument at high flow rates 38 Monitoring...

Страница 15: ...ing the exhaust trap bottle 63 Emptying the roughing pump exhaust liquid trap bottle 65 Gas ballasting the roughing pump 68 Gas ballasting a pump fitted with a screwdriver operated gas ballast valve 6...

Страница 16: ...ource assembly 106 Replacing the ESI probe tip 108 Replacing the ESI probe sample capillary 109 Replacing the ion block source heater 116 Replacing the source assembly seals 121 Removing the source en...

Страница 17: ...vacuum connections 158 Connecting the oil filled roughing pump 159 Making the electrical connections for a roughing pump with an external relay box 164 Making the electrical connections for a roughing...

Страница 18: ...2 71500126802IVD Rev C Connecting to the electricity source 187 C Materials of Construction and Compliant Solvents 189 Preventing contamination 190 Items exposed to solvent 190 Solvents used to prepar...

Страница 19: ...uding its controls and gas and plumbing connections Contents Topic Page Overview 20 Ionization techniques and source probes 24 Ion optics 25 MS operating modes 26 MS MS operating modes 27 Sample inlet...

Страница 20: ...eric pressure ionization API mass spectrometer Designed for routine UPLC MS MS analyses in quantitative and qualitative applications it can operate at fast acquisition speeds compatible with ultra per...

Страница 21: ...les simplified setup of the system for use in routine analytical and open access applications See Software and data system on page 1 22 The IntelliStart Fluidics system is built into the instrument It...

Страница 22: ...or Controlling Contamination in LC MS Systems part number 715001307 You can find the documents on http www waters com Software and data system The instrument is controlled by MassLynx mass spectromet...

Страница 23: ...Overview December 6 2012 71500126802IVD Rev C 23 the system and its modules The ACQUITY UPLC Console functions independently of the MassLynx software and does not recognize or control the data systems...

Страница 24: ...aerosol undergo a reduction in size solvent evaporation As solvent continues to evaporate the charge density increases until the droplet surfaces eject ions ion evaporation The ions can be singly or m...

Страница 25: ...o their mass to charge ratios 4 The mass separated ions pass into the T Wave collision cell where they either undergo collision induced dissociation CID or pass to the second quadrupole Any fragment i...

Страница 26: ...useful tool for instrument tuning and calibration before MS MS analysis and for fault diagnosis See the instrument online Help for further information on optimizing the TQ Detector for fast MS scannin...

Страница 27: ...e Typical applications Product ion mode is typically used for the following applications Method development for MRM screening studies MS MS operating modes Operating mode MS1 Collision cell MS2 Produc...

Страница 28: ...m product scan data by scanning for all the precursors of a common product ion Multiple reaction monitoring mode MRM mode is a highly selective MS MS equivalent of Selected Ion Recording SIR Because b...

Страница 29: ...al loss mode detects the loss of a specific neutral fragment or functional group from an unspecified precursor s The scans of MS1 and MS2 are synchronized When MS1 transmits a specific precursor ion M...

Страница 30: ...leak sensor stops system flow when its optical sensor detects about 1 5 mL of accumulated leaked liquid in its surrounding reservoir At the same time the ACQUITY UPLC Console displays an error message...

Страница 31: ...ACN 065444751 1 2 3 4 5 6 7 8 9 10 Anal og Notused St op Fl ow Swi t ch 2 G r ound G r ound O ut O ut 1 2 3 4 5 6 7 8 9 10 I nj ectS t ar t Event Swi t ch 3 Swi t ch 4 G r ound G r ound I n I n O ut...

Страница 32: ...ect or combined infusion to enable optimization at analytical flow rates The system incorporates a multi position valve with these attributes An input connection from an external LC column An input co...

Страница 33: ...development is automatically performed by the software IntelliStart Fluidics configuration requirements can be set in the system console You can edit the parameters frequency and extent of the automat...

Страница 34: ...1 Waters TQ Detector 34 December 6 2012 71500126802IVD Rev C...

Страница 35: ...for Operation This chapter describes how to start and shut down the instrument Contents Topic Page Starting the instrument 36 Preparing the IntelliStart Fluidics system 40 Rebooting the instrument 45...

Страница 36: ...d oil free with a purity of at least 95 Regulate the supply at 600 to 690 kPa 6 0 to 6 9 bar 90 to 100 psi For more information on connections see the figure on page 31 2 Ensure that the collision gas...

Страница 37: ...Operate LED remain off 6 Start the MassLynx software 7 Click IntelliStart in the MassLynx main window s lower left hand corner to launch IntelliStart Result The TQ Detector s console appears The instr...

Страница 38: ...Properties 5 Click OK Verifying the instrument s state of readiness When the instrument is in good operating condition the power and Operate LEDs show constant green You can view any error messages i...

Страница 39: ...t indicate its operational status Power LED The power LED to the top left hand side of the instrument s front panel indicates when the instrument is powered on or powered off Operate LED The Operate L...

Страница 40: ...lvent manifold drip tray Required material Chemical resistant powder free gloves To install the solvent manifold drip tray Install the solvent manifold drip tray as shown below Warning The solvent man...

Страница 41: ...stant powder free gloves To install the reservoir bottles 1 Remove the reservoir bottle caps 2 Screw the reservoir bottles onto the instrument as shown below To install low volume vials 1 If a standar...

Страница 42: ...t 3 Screw the low volume vial into the adaptor Diverter valve positions Column and syringe in home position after power up After power up the flow path between the column and waste is open The syringe...

Страница 43: ...the LC position the flow path between the LC and probe is open and the flow path between the syringe and waste is also open Infusion position Column LC LC Waste Probe Syringe Idle Reservoir A Reservoi...

Страница 44: ...position In the waste position both the LC flow and the infusion syringe flow are directed to waste The syringe mode can be only static or dispensing that is never drawing Column Combined LC Waste Pro...

Страница 45: ...ize carryover Rebooting the instrument Reboot the instrument when any of these conditions applies The Tune window fails to respond MassLynx software fails to initialize Immediately following a softwar...

Страница 46: ...nstances put the mass spectrometer in Standby mode See the online Help for details Emergency instrument shutdown To shut down the instrument in an emergency 1 Operate the power button on the front of...

Страница 47: ...71500126802IVD Rev C 47 3 ESI Operation This chapter describes how to prepare the instrument for ESI electrospray ionization Contents Topic Page Introduction 48 Installing the ESI probe 48 Removing t...

Страница 48: ...r instruction see page 24 Installing the ESI probe Probe compatibility warning label Required material Chemical resistant powder free gloves Warning On instruments bearing the probe compatibility warn...

Страница 49: ...ons ESI probe and source can be contaminated with biohazardous and or toxic materials Always wear chemical resistant powder free gloves while performing this procedure Warning To avoid electric shock...

Страница 50: ...acts on the ESI probe align with the probe adjuster assembly contacts and carefully slide the ESI probe into the hole in the probe adjuster assembly 5 Secure the ESI probe by tightening the 2 thumbscr...

Страница 51: ...different ID are supplied with the instrument Requirement If you are replacing the tubing supplied with the instrument you must minimize the length of the tube connecting the diverter valve to the ES...

Страница 52: ...probe to the probe adjuster assembly 8 Carefully remove the ESI probe from the probe adjuster assembly 9 If available fit the protective sleeve to the ESI probe tip 10 Close the instrument s access d...

Страница 53: ...rce 59 Operating the source isolation valve 60 Removing O rings and seals 62 Cleaning the instrument s exterior 63 Emptying the exhaust trap bottle 63 Emptying the roughing pump exhaust liquid trap bo...

Страница 54: ...ance Procedures 54 December 6 2012 71500126802IVD Rev C Maintaining the instrument air filters 131 Replacing the roughing pump oil 135 Replacing the roughing pump s oil demister element 138 Contents T...

Страница 55: ...rap bottle in the roughing pump exhaust line Check daily empty as required See page 65 Gas ballast the roughing pump ESI weekly See page 68 Inspect and adjust the roughing pump oil level Weekly See pa...

Страница 56: ...h you control the system The service agent software automatically and securely captures and sends information about the support needs of your system directly to Waters If you encounter a performance i...

Страница 57: ...d allow approximately 5 minutes to save the service profile Result A zip file containing your Connections INSIGHT profile is forwarded to Waters customer support for review Tip Saving a service profil...

Страница 58: ...de the instrument ensure that the instrument is in Standby mode before commencing any maintenance Warning The probe and source can be hot To avoid burn injuries take great care while working with thes...

Страница 59: ...urce To prepare the instrument 1 In the Instrument Console click Stop Flow to stop the LC flow or if column flow is required divert the LC flow to waste as follows a In the Instrument Console system t...

Страница 60: ...asing both spring clips and lowering the door towards you Warning The source components can be contaminated with biohazardous and or toxic materials Always wear chemical resistant powder free gloves w...

Страница 61: ...urce enclosure door and secure it with both spring clips 3 Close the instrument s access door Warning The source components can be contaminated with biohazardous and or toxic materials Always wear che...

Страница 62: ...removal kit To remove an O ring 1 Use the forked end of tool 1 to impale the O ring or seal 2 Pull the O ring or seal from its groove if necessary use tool 2 as an aid 3 Dispose of the O ring or seal...

Страница 63: ...xhaust trap bottle Check the exhaust trap bottle in the instrument exhaust line daily and empty it before it is completely full Instrument exhaust trap bottle Required materials Chemical resistant pow...

Страница 64: ...ion gas flow 8 In the ACQUITY UPLC Console click Start Flow to start the LC flow Warning The waste liquid in the exhaust trap bottle comprises ACQUITY UPLC solvents and analytes Always wear chemical r...

Страница 65: ...etector liquid to ensure that there are no leaks at the exhaust trap bottle cap Emptying the roughing pump exhaust liquid trap bottle Check the liquid trap bottle in the roughing pump exhaust line dai...

Страница 66: ...tor liquid To empty the exhaust liquid trap bottle 1 Close the source isolation valve see page 60 Warning The liquid in the roughing pump liquid trap bottle can be contaminated with analyte accumulate...

Страница 67: ...in the upright position 6 Open the source isolation valve see page 60 7 Use snoop or equivalent leak detector liquid to ensure that there are no leaks at the liquid trap bottle cap Warning The waste...

Страница 68: ...ballast the roughing pump when these conditions apply With ESI operation once a week When the roughing pump oil appears cloudy When the vacuum pressure is higher than normal When condensate forms in t...

Страница 69: ...osition 2 Run the pump for 30 to 60 minutes Tip It is normal for the roughing pump temperature to increase during ballasting To maintain an ambient temperature of 40 C 104 F where the pump is located...

Страница 70: ...tion 2 Run the pump for 30 to 60 minutes Tip It is normal for the roughing pump temperature to increase during ballasting To maintain an ambient temperature of 40 C 104 F where the pump is located ens...

Страница 71: ...at it is at or near the MAX level when the pump is not operating Tip The oil level in the sight glass is lower when the roughing pump is running than when it is stopped When the pump is running the oi...

Страница 72: ...rolled by the O ring groove depth in the plug Increased torque does not improve the plug seal it only makes the plug difficult to remove later 6 Start the instrument see page 36 Tips After you add oil...

Страница 73: ...led oil smells slightly Cleaning the source components Clean the sample cone and gas cone when these conditions apply The sample cone and gas cone are visibly fouled LC and sample related causes for d...

Страница 74: ...The source components can be contaminated with biohazardous and or toxic materials Always wear chemical resistant powder free gloves while performing this procedure Warning To avoid electric shock ens...

Страница 75: ...cone gas assembly connection tube 3 Grasp the cone gas assembly connection tube and use it as a lever to rotate the cone gas assembly 90 degrees moving the connection tube from the vertical to the hor...

Страница 76: ...2012 71500126802IVD Rev C 4 Slide the cone gas assembly out of the ion block assembly Caution Do not open the isolation valve at any time when the cone gas assembly has been removed from the ion bloc...

Страница 77: ...d with the instrument This tool is used in a similar manner to that fitted to the source enclosure door Warning The cone gas assembly can be contaminated with biohazardous and or toxic materials Alway...

Страница 78: ...ves Appropriately sized glass vessels in which to completely immerse components when cleaning Use only glassware not previously cleaned with surfactants HPLC grade or better methanol HPLC grade or bet...

Страница 79: ...taining water and then placing the vessels in the ultrasonic bath for 20 minutes b Displace the water by immersing the components in separate glass vessels containing methanol and then placing the ves...

Страница 80: ...present do as follows a Use the wash bottle containing 1 1 methanol water to rinse the component over the large beaker b Blow dry the component with inert oil free gas 7 Inspect each component for pe...

Страница 81: ...assembly 1 Carefully fit the sample cone into the gas cone 2 Fit a new O ring into the groove created between the sample cone and gas cone Caution To avoid recontaminating the cone gas assembly wear...

Страница 82: ...ardous and or toxic materials Always wear chemical resistant powder free gloves while performing this procedure Warning To avoid puncture wounds take great care while working with the source enclosure...

Страница 83: ...embly into the ion block assembly 3 Grasp the cone gas assembly connection tube and use it as a handle to rotate the cone gas assembly 90 degrees moving the connection tube from the horizontal to the...

Страница 84: ...3 Open the source enclosure door by releasing both spring clips and lowering the door towards you 4 Close the source isolation valve see page 60 5 Disconnect the PTFE tube from the cone gas assembly...

Страница 85: ...2 71500126802IVD Rev C 85 6 Use the 6 mm Allen wrench to unscrew and remove the 2 ion block assembly securing screws and associated washers 7 Remove the ion block assembly from the PEEK ion block supp...

Страница 86: ...ion block assembly 1 Ensure that the isolation valve is closed 2 Grasp the cone gas assembly connection tube and use it as a lever to rotate the cone gas assembly 90 degrees Warning The ion block asse...

Страница 87: ...en wrench to loosen the 4 ion block cover plate captive securing screws 5 Remove the ion block cover plate Caution To ensure correct operation of the ion block assembly after reassembly the gas cone p...

Страница 88: ...to carefully remove the isolation valve O ring see page 62 8 Dispose of the isolation valve O ring in accordance with local environmental regulations Warning The isolation valve O ring can be contami...

Страница 89: ...the 2 set screws securing the heater cartridge to the PEEK terminal block are fitted use the 1 5 mm Allen wrench to loosen them 10 Use the 2 5 mm Allen wrench to loosen the captive PEEK terminal block...

Страница 90: ...eedle nose pliers to gently grasp the heat shrink tubing on the heater cartridge assembly and slide it and the PEEK terminal block out of the ion block Caution To avoid damaging the heater cartridge a...

Страница 91: ...refully remove the cover seal from the ion block see also page 62 14 Dispose of the cover seal in accordance with local environmental regulations Warning The cover seal can be contaminated with biohaz...

Страница 92: ...king plug and associated seal 16 Dispose of the blanking plug seal in accordance with local environmental regulations Warning The blanking plug seal can be contaminated with biohazardous and or toxic...

Страница 93: ...w 18 Remove the extraction cone retainer and extraction cone from the ion block Caution Take great care not to damage the extraction cone aperture when removing the extraction cone from the ion block...

Страница 94: ...ls Chemical resistant powder free gloves Appropriately sized glass vessels in which to completely immerse components when cleaning Use only glassware not previously cleaned with surfactants HPLC grade...

Страница 95: ...ng the vessels in the ultrasonic bath for 10 minutes 4 Carefully remove the components from the vessels and blow dry them using inert oil free gas 5 Inspect each component for persisting contamination...

Страница 96: ...r To assemble the ion block assembly 1 Fit the extraction cone seal to the ion block 2 Fit the extraction cone retainer to the extraction cone 3 Fit the extraction cone retainer and extraction cone to...

Страница 97: ...the cover seal and extraction cone seal are free from dirt and debris Tip If contamination is present use 1 1 methanol water applied to a lint free cloth to carefully clean the grooves 11 Fit a new co...

Страница 98: ...lock from the PEEK ion block support on the instrument 19 Fit the ion block cover plate to the ion block assembly and then use the 2 5 mm Allen wrench to tighten the 4 ion block cover plate captive se...

Страница 99: ...rce enclosure door by releasing both spring clips and lowering the door towards you 3 Remove the side flange from the source enclosure Rationale This step allows you to see the ion block assembly as y...

Страница 100: ...osition during installation If this is the case remove the ion block assembly refit the extraction cone and reinstall the ion block assembly on the PEEK ion block support 7 When you are satisfied with...

Страница 101: ...l resistant powder free gloves 3 mm Allen wrench O ring removal kit To remove the ion block assembly ion block support and hexapole 1 Remove the ion block assembly from the PEEK ion block support see...

Страница 102: ...environmental regulations 6 Carefully grasp the source hexapole assembly and remove it from the adaptor housing Warning The O rings can be contaminated with biohazardous and or toxic materials Ensure...

Страница 103: ...methanol Ultrasonic bath Source of oil free inert gas nitrogen or helium for drying air drying optional HPLC grade or better 1 1 methanol water Wash bottle containing HPLC grade or better 1 1 methanol...

Страница 104: ...mersed completely 5 Place the vessel in the ultrasonic bath for 30 minutes 6 Carefully remove the hexapole assembly from the vessel and blow dry it using inert oil free gas Caution To avoid vibration...

Страница 105: ...rsisting contamination If contamination is present do as follows a Use the wash bottle containing methanol to rinse the source hexapole assembly over the large beaker b Blow dry the hexapole assembly...

Страница 106: ...rt O rings are free from dirt and debris Tip If contamination is present use 1 1 methanol water applied to a lint free cloth to carefully clean the grooves 3 Fit new O rings to the PEEK ion block supp...

Страница 107: ...and tighten the 3 PEEK ion block support securing screws Tip Look through the PEEK ion block support as you tighten the securing screws Ensure that neither hexapole spring buckles and touches a hexapo...

Страница 108: ...d tighten the new probe tip to the probe 5 Adjust the probe tip so that the fully extended capillary when the probe nebulizer adjuster knob is fully screwed down protrudes by approximately 1 to 1 5 mm...

Страница 109: ...h wrench 7 16 inch wrench Flat blade screwdriver Needle nose pliers LC pump HPLC grade or better 1 1 acetonitrile water To replace the capillary 1 Remove the probe from the source see page 52 Warning...

Страница 110: ...s 110 December 6 2012 71500126802IVD Rev C 2 Use the screwdriver to remove the 2 probe end cover retaining screws 3 Use the 1 5 mm Allen wrench to loosen the 3 set screws on the LC PEEK union 2 retain...

Страница 111: ...Replacing the ESI probe sample capillary December 6 2012 71500126802IVD Rev C 111 4 Remove the end cover 5 Use the 6 mm wrench to remove the probe tip End cover TP03007 Probe tip...

Страница 112: ...ry PTFE liner ferrule assembly and seal in accordance with local environmental regulations 10 Use the needle nose pliers to remove the conductive sleeve from the inner bore of the probe assembly fitti...

Страница 113: ...end of the LC union 15 Lightly screw the coupling into the LC union so that it does not restrict movement of the capillary 16 Slide a compression screw and ferrule onto a piece of 1 16 0 0025 bore 5M...

Страница 114: ...n and then repeat the leak test 21 When performing the leak test check the backpressure on the LC pump which will be relatively high if the capillary is blocked If this is the case replace the capilla...

Страница 115: ...e the probe tip and then screw down until the capillary protrudes approximately 0 5 mm from the end of the tip 27 Fit the probe end cover 28 Fit and tighten the two probe end cover retaining screws 29...

Страница 116: ...lock source heater 1 Remove the ion block assembly from the instrument see page 84 2 Ensure that the isolation valve is closed Warning The capillary and ferrule can be contaminated with biohazardous a...

Страница 117: ...tive screws securing the ion block cover plate 4 Remove the ion block cover plate 5 Use the 1 5 mm Allen wrench to remove the 2 screws securing the heater wires to the PEEK terminal block TP02649 Ion...

Страница 118: ...nose pliers to carefully swing the ring terminal tags out of the terminal block 7 If the 2 set screws securing the heater cartridge to the PEEK terminal block are fitted use the 1 5 mm Allen wrench to...

Страница 119: ...PEEK terminal block are fitted use the 1 5 mm Allen wrench to fit and tighten them 12 Use the needle nose pliers to position the 2 heater wire ring tags fully down on the PEEK block terminals Caution...

Страница 120: ...fit and tighten the 2 screws securing the heater wires to the PEEK terminal block 14 Fit the ion block cover plate to the ion block assembly and then use the 2 5 mm Allen wrench to tighten the 4 capti...

Страница 121: ...Pressure Test Unit Operator s Guide Removing the source enclosure from the instrument Required materials Chemical resistant powder free gloves 5 mm Allen wrench To remove the source enclosure 1 Vent a...

Страница 122: ...connection on the front panel 6 Disconnect the 2 pin connector at the rear of the source enclosure 7 Use the 5 mm Allen wrench to loosen the 3 captive source enclosure securing screws 8 Remove the sou...

Страница 123: ...o the source enclosure 2 Remove the probe adjuster assembly from the source enclosure 3 Unscrew and remove the 3 thumbscrews and door spring clip that secure the source enclosure side flange to the so...

Страница 124: ...y 1 Use the O ring removal kit to carefully remove the probe adjuster assembly probe seal from the probe adjuster assembly see page 62 2 Use the O ring removal kit to carefully remove the following se...

Страница 125: ...flange seal 3 Open the source enclosure door by releasing both spring clips and lowering the door towards you Caution Do not apply any downward force to the source enclosure door while the door is op...

Страница 126: ...enclosure door 7 Remove the glass from the source enclosure door 8 Use the O ring removal kit to carefully remove the door glass seal from the source enclosure door 9 Dispose of all the seals in accor...

Страница 127: ...losure door 5 Fit and tighten the 4 screws securing the glass retaining clips to the source enclosure door 6 Fit the new source enclosure door seal to the source enclosure door 7 Close the source encl...

Страница 128: ...osure 1 Fit the source enclosure side flange to the source enclosure 2 Fit and tighten the 3 thumbscrews that secure the source enclosure side flange to the source enclosure 3 Fit the probe adjuster a...

Страница 129: ...Connect the PTFE tubing to the desolvation gas connection at the instrument s front panel 5 Connect the probe s electrical connection at the instrument s front panel 6 Close the source enclosure door...

Страница 130: ...36 9 Perform a source pressure test as described in the Waters Micromass Source Pressure Test Unit Operator s Guide Warning To confirm the integrity of the source exhaust system a source pressure test...

Страница 131: ...strument s lower bezel Required materials T10 TORX driver Vacuum cleaner To clean the air filter inside the instrument s lower bezel 1 Use the T10 TORX driver to remove the 5 screws that secure the lo...

Страница 132: ...air filter by vacuuming replace it with a new filter Required materials T10 TORX driver TQ Detector lower bezel air filter To replace the air filter inside the lower bezel 1 Remove the 5 screws that s...

Страница 133: ...he T10 TORX driver to fit and tighten the 4 screws securing the air filter and frame to the inside of the lower bezel 6 Reinstall the lower bezel Cleaning the air filter behind the source probe Requir...

Страница 134: ...1500126802IVD Rev C Air filter tab 3 Use a vacuum cleaner to clean the air filter Air filter removed from behind the source probe 4 Reinstall the air filter 5 Fit the probe to the source see page 48 R...

Страница 135: ...he figure on page 134 3 Remove the air filter from the air filter frame and dispose of it Removing air filter from frame 4 Insert the new air filter in the air filter frame 5 Reinstall the air filter...

Страница 136: ...d shut down the instrument see the mass spectrometer s online Help for details 3 Allow the roughing pump to cool 4 Place the container for used oil under the pump s drain plug 5 Use the 8 mm Allen wre...

Страница 137: ...is controlled by the O ring groove depth in the plug Increased torque does not improve the plug seal it only makes the plug difficult to remove later 11 Using the funnel pour all the oil from the 1 L...

Страница 138: ...plug will run down and drip off the pump once the pump reaches operating temperature When the pump begins to run at normal operating temperature spilled oil smells slightly Replacing the roughing pump...

Страница 139: ...bolts securing the exhaust flange to the roughing pump Warning The pump oil can be contaminated with analyte accumulated during normal operation Always wear chemical resistant powder free gloves when...

Страница 140: ...C 4 Using both hands carefully remove the exhaust flange and oil demister element from the roughing pump 5 Use the 10 mm wrench to remove the nut that secures the oil demister element to the exhaust...

Страница 141: ...emove the exhaust flange from the oil demister element 7 Remove the spring from the oil demister element 8 Dispose of the oil demister element in accordance with local environmental regulations Warnin...

Страница 142: ...o fit and tighten the nut that secures the oil demister element to the exhaust flange Warning The pump oil can be contaminated with analyte accumulated during normal operation Always wear chemical res...

Страница 143: ...oth hands carefully fit the oil demister element and exhaust flange to the roughing pump 5 Use the 6 mm Allen wrench to fit the 4 bolts securing the exhaust flange to the roughing pump 6 Start the ins...

Страница 144: ...4 Maintenance Procedures 144 December 6 2012 71500126802IVD Rev C...

Страница 145: ...responding user guides also include the hazard symbols with accompanying text statements describing the hazards and telling you how to avoid them This appendix presents all the safety symbols and stat...

Страница 146: ...nt component Such risks include burn injuries electric shocks ultraviolet radiation exposures and others When the following symbols appear in a manual s narratives or procedures their accompanying tex...

Страница 147: ...n pose a serious health threat Warning Risk of tipping Warning Risk of explosion Warning Pressurized nonmetallic or polymer tubing can burst Observe these precautions when working around such tubing W...

Страница 148: ...that enclosed space Ensure that the nitrogen supply pressure never falls below 690 kPa 6 9 bar 100 psi during an analysis in which flammable solvents are used Also ensure a gas fail connection is conn...

Страница 149: ...ctivated microorganisms and other biological materials To avoid infection with these agents assume that all biological fluids are infectious observe Good Laboratory Practices and consult your organiza...

Страница 150: ...ementation peut annuler le droit de l utilisateur exploiter l quipement Achtung Jedwede nderungen oder Modifikationen an dem Ger t ohne die ausdr ckliche Genehmigung der f r die ordnungsgem e Funktion...

Страница 151: ...nez toute flamme se trouvant proximit de l instrument Evitez d utiliser des tubes s v rement d form s ou endommag s Evitez d utiliser des tubes non m talliques avec du t trahydrofurane THF ou de l aci...

Страница 152: ...do provocano rigonfiamenti nei tubi non metallici riducendo notevolmente la pressione di rottura dei tubi stessi Advertencia se recomienda precauci n cuando se trabaje con tubos de pol mero sometidos...

Страница 153: ...r le mat riel risque d tre d fectueuses Vorsicht Der Benutzer wird darauf aufmerksam gemacht dass bei unsachgem er Verwenddung des Ger tes die eingebauten Sicherheitseinrichtungen unter Umst nden nich...

Страница 154: ...nd handling symbols Electrical symbols These can appear in instrument user manuals and on the instrument s front or rear panels Electrical power on Electrical power off Standby Direct current Alternat...

Страница 155: ...r 6 2012 71500126802IVD Rev C 155 Handling symbols These handling symbols and their associated text can appear on labels affixed to the outer packaging of Waters instrument and component shipments Kee...

Страница 156: ...A Safety Advisories 156 December 6 2012 71500126802IVD Rev C...

Страница 157: ...t or remove it from service contact Waters Technical Service for recommended cleaning flushing and packaging procedures See page 56 Contents Topic Page External wiring and vacuum connections 158 Conne...

Страница 158: ...d G r ound O ut O ut 1 2 3 4 5 6 7 8 9 10 I nj ectS t ar t Event Swi t ch 3 Swi t ch 4 G r ound G r ound I n I n O ut O ut O ut RS 232 RS 232 ETHERNET ETHERNET API Gas API Gas COLLISION GAS COLLISION...

Страница 159: ...p bottle included in the startup kit Elbows included in the Waters Rough Pump Connect Kit NW25 tee included in the startup kit NW25 center rings included in the startup kit NW25 clamps included in the...

Страница 160: ...t powder free gloves when performing this procedure Caution To ensure correct operation of the roughing pump the pump must be installed within 1 degree of horizontal The area where the roughing pump i...

Страница 161: ...to install the clamps 5 Connect the opposite ends of the two lengths of vacuum hose in step 4 to the two 1 inch OD straight vacuum ports on the instrument s rear panel Secure the hose ends with clamps...

Страница 162: ...kit 8 Connect a length of 12 7 mm clear PVC exhaust tubing to an elbow and connect the elbow to the other fitting on the exhaust trap bottle The exit line of the exhaust trap bottle can be at any elev...

Страница 163: ...x see page 164 If your roughing pump does not have an external relay box see page 165 Caution The instrument requires two separate exhaust systems one for nitrogen the other for the roughing pump Vent...

Страница 164: ...connector on the instrument s rear panel 01757 U S A V 200 240 V 200 240 Hz 50 60 Hz 50 60 V VA 900 ACN 065444751 1 2 3 4 5 6 7 8 9 10 Anal og Notused St op Fl ow Sw i t ch 2 G r ound G r ound O ut O...

Страница 165: ...ing pump d c connector to the pump connector on the instrument s rear panel 01757 U S A V 200 240 V 200 240 Hz 50 60 Hz 50 60 V VA 900 ACN 065444751 1 2 3 4 5 6 7 8 9 10 Anal og Notused St op Fl ow Sw...

Страница 166: ...o minimize operating noise unless the pump is installed in a cabinet that includes sound dampening and ventilation Required materials Chemical resistant powder free gloves 7 mm nut driver 8 mm Allen w...

Страница 167: ...mp Connect Kit 1 inch ID vacuum hose included in the Waters Rough Pump Connect Kit Isolation valve included in the Alcatel pump kit External silencer included in the Alcatel pump kit Noise reduction c...

Страница 168: ...B External Connections 168 December 6 2012 71500126802IVD Rev C 6 Install the noise reduction cover TP02996 DN 25 nipple DN 40 nipple Clamps...

Страница 169: ...ect operation of the roughing pump the pump must be installed within 1 degree of horizontal The area where the roughing pump is located must have an ambient temperature of 15 to 40 C 59 to 104 F To en...

Страница 170: ...all the clamps 4 Connect the opposite ends of the two lengths of vacuum hose in step 3 to the two 1 inch OD straight vacuum ports on the instrument s rear panel Secure the hose ends by installing 2 cl...

Страница 171: ...nection with a clamp 8 Connect an approximately 300 mm 12 inch length of 12 7 mm clear PVC exhaust tubing to the DN25 reducing nipple securing the tubing with a hose clamp 9 Connect an elbow to the ot...

Страница 172: ...ap bottle The exit line of the exhaust trap bottle can be at any elevation 11 Secure the exhaust trap bottle in a conspicuous location 12 Route the open end of the exhaust tubing to a suitable exhaust...

Страница 173: ...ct the power cable and power on the instrument 01757 U S A V 200 240 V 200 240 Hz 50 60 Hz 50 60 V VA 900 ACN 065444751 1 2 3 4 5 6 7 8 9 10 Anal og Notused St op Fl ow Swi t ch 2 G r ound G r ound O...

Страница 174: ...sharp knife to cut a 3 8 to 5 0 cm 1 5 to 2 inch length of 6 mm PTFE tubing 2 Connect this piece of tubing to one end of the nitrogen supply in line filter 3 Connect the remaining length of the 6 mm...

Страница 175: ...50 60 V VA 900 ACN 065444751 1 2 3 4 5 6 7 8 9 10 Anal og Notused St op Fl ow Swi t ch 2 G r ound G r ound O ut O ut 1 2 3 4 5 6 7 8 9 10 I nj ectS t ar t Event Swi t ch 3 Swi t ch 4 G r ound G r ound...

Страница 176: ...the 1 8 inch Swagelok nut and ferrule to connect the 1 8 inch stainless steel tube to the collision cell gas inlet on the rear of the instrument see the figure on page 175 2 Use the wrench to tighten...

Страница 177: ...m vacuum of 0 20 kPa 2 mbar 0 03 psi below atmospheric pressure negative pressure The exhaust connections can be contaminated with biohazardous and or toxic materials Always wear chemical resistant po...

Страница 178: ...strument exhaust and laboratory exhaust system line connections Exhaust trap bottle Warning To confirm the integrity of the source exhaust system the following leak test must be performed Caution To a...

Страница 179: ...ide a drain line over the barbed fitting of the drain located at the bottom of the instrument Warning The waste line and connection can be contaminated with biologically hazardous materials Always wea...

Страница 180: ...nd is above the surface of the waste solvent Positioning of drain tube Warning To prevent leakage of biologically hazardous materials ensure that the drain line does not crimp or bend A crimp or bend...

Страница 181: ...ect the workstation 1 Connect the monitor to the PC 2 Connect one end of the shielded network cable to the appropriate port on the rear panel of the instrument 3 Connect the other end of the shielded...

Страница 182: ...anel includes two removable connectors that hold the screw terminals for I O signals These connectors are keyed so that they can receive a signal cable inserted only one way Warning To avoid electric...

Страница 183: ...3 4 5 6 7 8 9 10 Analog Out Analog Out Ground Not Used Not Used Stop Flow Out Stop Flow Out Ground Switch 2 Out Switch 2 Out Connector I 1 2 3 4 5 6 7 8 9 10 Inject Start In Inject Start In Ground Ev...

Страница 184: ...al connections Description Analog Out Used for analog chart output functionality The output voltage range is 0 to 2 V The resolution of the voltage output is 12 bits Stop Flow Out Used to stop the sol...

Страница 185: ...egative leads of the signal cable to the connector 3 Slide the clamp with the bend facing down into the protective shield 4 Insert the clamp and shield with the bend facing down into the connection co...

Страница 186: ...the signal cable into the connection cover and position the clamp over the cable leads Tighten the clamp into place with the second self tapping screw 6 Place the second connection cover over the fir...

Страница 187: ...d of the power cord to the receptacle on the rear panel of the instrument 2 Connect the male end of the instrument power cord to the 200 to 240 V AC wall outlet prepared as described in the instrument...

Страница 188: ...B External Connections 188 December 6 2012 71500126802IVD Rev C...

Страница 189: ...mpliant Solvents Warning To confirm the integrity of the source exhaust system you must address any safety issues raised by the contents of this Appendix Contents Topic Page Preventing contamination 1...

Страница 190: ...used with these items See page 191 for details about the most common ingredients used to prepare mobile phases Items exposed to solvent Item Material Autotune reservoirs High density polyethylene Cor...

Страница 191: ...ile Formic acid 0 1 Acetic acid 0 1 Trifluoroacetic acid 0 1 Ammonium acetate 10 mM Ammonium formate 10 mM These solvents are not expected to cause any problems with the materials identified in the ta...

Страница 192: ...C Materials of Construction and Compliant Solvents 192 December 6 2012 71500126802IVD Rev C...