WL100
Manual Page 39 Revision: 1‐14‐2019
8.
Once flushed, install the filters. Following the flow direction on the filter.
NOTE:
Filters should not be flushed prior to 24 hours before installation to limit Microbial
Growth.
9.
Connect
WL100 Water Treatment System
to power.
CAUTION!
NEVER TURN ON HEATER BEFORE FILLING HOT TANK.
Red Compressor/Heater Switch must be in the O=OFF position while the Hot
Tank is empty. Damage could occur within one minute and the overload (high
limit) will require manual reset if Heater is turned on with an empty Hot Tank.
Flushing the Sanitizer from the Machine
10.
Place a pitcher, catch basin, or other container under the faucet of the
WL100 Water Treatment
System
.
11.
Flush the Cold Tank. Run several gallons of water through the faucet by dispensing cold water to
dilute and remove the sanitizer from the cold circuit. Chlorine test strips can be used to evaluate
the water.
12.
Once the sanitizer odor/taste has been flushed out of the cold side of the machine the
sanitization process for the Cold Circuit is now complete.
Fill the Hot Tank
13.
Press the Hot Water Select Button, followed by the Main Dispensing Button to fill the Hot Tank.
Water will dispense from the faucet once the Hot Tank is full. Flush until water is clear.
WARNING!
HOT CIRCUIT IS NOT SANITIZED.
Water in the Hot Circuit is not sanitary until the temperature exceeds 77°C (171°F) for at least
5 minutes.
Compressor Test
14.
Switch Red Compressor / Heater to on
I=ON position
. Always ensure Tanks are
full of water before turning on the Heater or the Overload (High Limit) opens and
require manual reset. If the Wire Condenser at back of the
WL100 Water
Treatment System
is warm, the refrigeration system is working.
15.
Once the machine reaches its target temperature, the Compressor will shut off. Draw a glass of
cold water and verify it has chilled to proper temperature.