Message / alert
DESCRIPTION AND ACTIONS
Air was detected in the chamber: a vacuum leak is suspected. The
cycle was completed but a vacuum test is required.
Run a vacuum test.
Call for service if an anomaly is detected.
Error when saving the files in the network.
Check the settings, the network resource and the
memory space available on it.
Some cycle reports couldn't be saved in the network, because the
sterilizer memory is limited (see chapter 7 – CYCLE HISTORY).
No action possible. The cycles were overwritten in the
sterilizer database before being saved, and there is no
way to recover them.
These messages advise that one consumable must be replaced.
Replace the consumable and CONFIRM to reset the
counter (See Chapter 8 – MAINTENANCE).
If you don't replace the consumable, press NO.
In this case, you may still use the sterilizer but the
message will appear again after some cycles.
WARNING!
Operating the sterilizer with expired
consumables could be dangerous and
could damage the sterilizer.
This message advises you that the target of 4000 cycles has been
achieved and the relevant maintenance step is to be performed.
Call technical service for the 4000 cycle maintenance.
71
Messages and alerts
NOTE: for any error not listed in this table, call technical service.