69
ERROR CODE
DESCRIPTION AND ACTIONS
18X
Chamber filter clogged. Remove and clean the chamber filter. See error 13X-16X
Repeat the cycle.
If the problem persists, call
service.
19X
Clean the door seal and the chamber face side.
2xx
Switch the sterilizer OFF and ON.
Wait for the chamber to cool down. Reset the thermal overload (see next pages).
3xx
Check the door seal. Clean or replace it if necessary.
Clean the chamber face side.
Clean the chamber filter.
Check if chamber filter is properly locked in the cap.
Check the load does not exceed the MAXIMUM WEIGHT LIMITS.
4xx
Clean water error (bad quality or low water level). Drain and refill the clean water tank.
5xx
Check if there are hurdles on the door locking area (chamber rack, loads, objects, …).
Check the door seal (wrong placed).
Check the door can move freely without touching the trays or the load when closing
Switch the sterilizer OFF and ON.
990
The cycle has been aborted by the user.
Re-process the load.
Error table (continued)
NOTE: for any error not listed in this table, call technical service.