During the sterilization cycle, the sterilizer is continuously monitored by a sophisticated control system.
If an anomaly is detected, the cycle is aborted automatically, and the sterilizer starts a reset phase.
Do not switch off the sterilizer during the reset phase: it takes some minutes to reset
the system and reach safe conditions in the sterilizer chamber.
Water could be present in the chamber when opening the door: prevent spilling (e.g., place a towel below the chamber door).
At the end of the reset phase, you may open the door.
- the message LOAD NOT STERILE means that the load is not
sterile. Do not use items on patients!
- the message DRYING INTERRUPTED means that the load
might be wet. Wet items are for immediate use only!
9. Troubleshooting, Errors, Messages and Alerts
The following information is displayed:
Error symbol
Error code
Warning message:
Wait message
You may display the INFO screen (see RUNNING A STERILIZATION CYCLE) by touching the INFO icon.