63
400-cycle maintenance
Replacing the bacteriological filter
Open the service door.
Unscrew the bacteriological filter by hand (counter-clockwise).
Screw on the new bacteriological filter (clockwise) and tighten it snug.
Replacing the dust filter
Pull out the dust filter from underneath the sterilizer.
Detach the used filter from the handle.
Attach the new filter to the handle.
Slide the filter back into its original position.
Remember to reset the counter after replacement
(see RESETTING THE MAINTENANCE COUNTERS).
Remember to reset the counter after replacement
(see RESETTING THE MAINTENANCE COUNTERS).
See video animation directly on Lisa display.
Also on WH.com FAQ videos.
See video animation directly on Lisa display.
Also on WH.com FAQ videos.