Maintenance
LINA | Instructions for Use | RIK-1 Med ENG Rev03 | 11/05/2020 | © 2020 W&H Sterilization Srl
71
5
Insert the new seal and press first up and down.
6
Press left and right and then the entire seal circumference to ensure its perfect
placement.
Notice: A steam discharge can damage the sterilizer plastic parts. Ensure the
seal does not stick out.
7
Wipe any residual water and run a Vacuum and Helix test to check for perfect
tightness of the seal. See "Vacuum test" on page 41 and "Helix test" on page 39.