Sterilizer tests
42
LINA | Instructions for Use | RIK-1 Med ENG Rev03 | 11/05/2020 | © 2020 W&H Sterilization Srl
CARRY OUT THE TEST
1
Empty the sterilizer chamber to ensure no load is present.
2
Close the chamber door and ensure the sterilizer chamber is completely dry and cold to avoid any false negative outcome.
3
On the homepage tap Vacuum test.
4
Tap
and enter your credentials if required: the chamber door locks.
5
Wait until the end of the test and tap OPEN: the chamber door unlocks.
6
Enter your credentials if required: a message informs if the test passed or failed. If the test failed, see "What to do when the test failed"
below
WHAT TO DO WHEN THE TEST FAILED
1
Check, clean or replace the door gasket.
2
Clean the chamber face side and the chamber filter.
3
Repeat the Vacuum test. See "Carry out the test" above.
4
If the test fails repeatedly, call technical service.