Procedure
To activate the remote storage and set the necessary parameters
do the following:
1
On the homepage tap
>
>
>
2
Tap
to enable the remote data storage: the first four
fields in the page and the check box turn dark grey.
3
In Path enter the name of the shared folder followed by the sub-
folder name, if any, where to save reports. Do not enter the full
path.
Note: The folder name must include letters and numbers only. Do not
use other characters like space-bar, slash, accent, etc.
4
Enter the host name or the IP address: if the data are complete,
the fields highlight.
5
Not mandatory. Enter the domain name.
6
Tap
to require the authentication credentials to access the
remote storage folder and enter the user name and password.
7
Tap
to save.
8
Tap
to go back to the previous page.
9
To check if the parameters entered are valid, see "Test the data
storage (optional)" below.
TEST THE DATA STORAGE (OPTIONAL)
Note: The test function is available only if the remote data storage is
enabled. See "Set the remote folder for saving the reports (optional)" on the
previous page.
1
On the homepage tap
>
>
2
Tap
: a sequence of tests is automatically performed.
3
In case a test fails, check the relevant settings and tap
to
repeat the test sequence, if the error persists call your IT
manager.
4
Tap
to go back to the previous page.
SAVE ALL THE CYCLE REPORTS IN A REMOTE FOLDER (OPTIONAL)
Note: The save all function is available only if the remote data storage is
enabled. See "Set the remote folder for saving the reports (optional)" on the
previous page.
Only the last 400 cycles in HTML and 100 cycles in SCL in the
sterilizer memory can be saved in the remote folder.
1
On the homepage tap
>
>
>
2
Tap
to start the remote saving.