Sterilizer tests
LINA | Instructions for Use | RIK-1 Med ENG Rev03 | 11/05/2020 | © 2020 W&H Sterilization Srl
37
PURPOSE OF THE TEST
The test is used to validate the sterilizer performance for textile load sterilization.
DESCRIPTION
It consists of several sheets of paper wrapped in a small packet with a chemical heat-sensitive indicator sheet in the middle. The colour
assumed by this indicator sheet at the end of the sterilization cycle gives the result of the test.
CARRY OUT THE TEST (IN ACCORDANCE WITH EN13060)
1
Empty the sterilizer chamber to ensure no load is present. Remove all trays from
the sterilization chamber, except the lowest one.
2
Place the Bowie and Dick test pack in the center of a tray in the lowest rack
position and close the chamber door.
3
On the homepage tap Helix-B&D test.
4
To set the duration of the Plateau/Sterilization phase and other settings, tap
.
5
Tap
and enter your credentials if required: the chamber door locks.
6
Wait until the end of the test and tap OPEN: the chamber door unlocks.
7
Enter your credentials if required.
8
Open the chamber door, extract the tray using the tray holder and take the test
pack. The test pack can be wet outside.
CAUTION! Risk of burns. The test pack is very hot at the end of the cycle.
Wear appropriate PPE (e.g. gloves).
9
Remove the indicator sheet from the center of the test pack and check the
change in colour. See "Interpret the test result" on the next page