Sterilizer tests
LINA | Instructions for Use | RIK-1 Med ENG Rev03 | 11/05/2020 | © 2020 W&H Sterilization Srl
41
WHAT TO DO NEXT
Compile the Helix test documentation form to trace the effectiveness of the sterilization cycle during the whole lifespan of your sterilizer. See
"Helix test documentation form" on page 110.
Vacuum test
CAUTION! Follow local/national guidelines on the frequency of testing.
Notice: If a drainage period of the S Naked 134 cycle is still operating, wait the drainage to be finished and the sterilizer to be both cold and dry.
Otherwise, a false negative outcome could occur.
PURPOSE OF THE TEST
The test is used to validate the sterilizer performance in terms of:
n
efficiency of the vacuum pump
n
tightness of the pneumatic circuit
DESCRIPTION
It consists of a vacuum phase, followed by a stabilization period of 5 minutes and a testing period of 10 minutes. The internal pressure is
monitored during the testing period. The pressure rise must be less than 0.013 bar (0.19 psi).